AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders; Product Usage: The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.
Enforcement
- Recall Number
- Z-2140-2012
- Event ID
- 62558
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Baxter Healthcare Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 15, 2012
- Initiation Date
- July 13, 2012
- Classification Date
- August 8, 2012
- Termination Date
- August 11, 2017
- Address
- Rt. 120 & Wilson Rd, Round Lake, IL, 60073, United States
Description
AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders; Product Usage: The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.
After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
Product code 2M8075, all serial numbers
Worldwide Distribution - USA - Nationwide including Puerto Rico, and internationally to Brazil, Canada, Colombia, Costa Rica, Czech Republic, El Salvador, Egypt, Germany, Guatemala, Honduras, Luxembourg, Mexico, Panama, Saudi Arabia, Singapore, Switzerland, Taiwan, Trinidad, Turkey, United Arab Emirates and United Kingdom.
840 units