FDA Enforcement Class II Ongoing

Aeos Robotic Digital Microscope, Product Code: PV010

Recall: Z-2139-2024 · Reported June 19, 2024

Enforcement

Recall Number
Z-2139-2024
Event ID
94623
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Aesculap Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
June 19, 2024
Initiation Date
May 13, 2024
Classification Date
June 12, 2024
Address
3773 Corporate Pkwy, N/A, Center Valley, PA, 18034-8217, United States

Description

Aeos Robotic Digital Microscope, Product Code: PV010

Reason

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Code Info

Product Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063

Distribution

US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.

Quantity

19 units