ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
Enforcement
- Recall Number
- Z-2139-2020
- Event ID
- 85390
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Radiometer Medical ApS
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- June 3, 2020
- Initiation Date
- November 21, 2019
- Classification Date
- May 28, 2020
- Termination Date
- November 8, 2024
- Address
- Akandevej 21, N/A, Bronshoj, N/A, N/A, Denmark
Description
ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
ABL90 FLEX Model no. 393-090: R0659 - R0977 and Spare Part Hemolyzer Units (902-980): R0526 - R0941
Worldwide distribution - US Nationwide. Also distributed OUS.
1233 units