FDA Enforcement Class II Ongoing

Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Recall: Z-2136-2021 · Reported August 4, 2021

Enforcement

Recall Number
Z-2136-2021
Event ID
88158
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medtronic Vascular, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 4, 2021
Initiation Date
June 7, 2021
Classification Date
July 23, 2021
Address
3850 Brickway Blvd, N/A, Santa Rosa, CA, 95403-8223, United States

Description

Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Reason

During product complaint investigation, it was determined that during stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover.

Code Info

Endurant IIs Stent Graft System: Model (REF):ESBF3214C103E GTIN: 00643169439962 Serial Numbers: V29780773; V29780771; V29780759; Model (REF): ESBF3614C103EE GTIN: 00643169780460 Serial Numbers: V29781932; V29781933; V29781934; V29781935; V29781936; V29781937; V29781938; V29781939; V29781940; V29781941; V29781942; V29781943; V29781944; V29781945; V29781946; V29781947; V29781948; V29781951;

Distribution

U.S.: GA, MA, MD, NY, SC, TX, WA, and WV O.U.S.: Argentina, Australia, Germany, Greece, Israel, Italy, Netherlands, Poland, South Africa, Spain, Switzerland and United Kingdom.

Quantity

21 devices