FDA Enforcement Class II Terminated

Y-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-C-01001); b) Large (003-C-02001)

Recall: Z-2135-2018 · Reported June 13, 2018

Enforcement

Recall Number
Z-2135-2018
Event ID
79819
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Orthofix, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 13, 2018
Initiation Date
April 6, 2018
Classification Date
June 7, 2018
Termination Date
June 30, 2021
Address
3451 Plano Pkwy, N/A, Lewisville, TX, 75056-9453, United States

Description

Y-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-C-01001); b) Large (003-C-02001)

Reason

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Code Info

(UDI)/Lot: a) (18056099647786) B1102163; (b (18056099647793) B1106912

Distribution

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Quantity

36