FDA Enforcement
Class II
Ongoing
SIGNA PET/MR
Recall: Z-2133-2025
·
Reported July 30, 2025
Enforcement
- Recall Number
- Z-2133-2025
- Event ID
- 97030
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- GE Medical Systems, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 30, 2025
- Initiation Date
- May 23, 2025
- Classification Date
- July 18, 2025
- Address
- 3200 N Grandview Blvd, N/A, Waukesha, WI, 53188-1693, United States
Description
SIGNA PET/MR
Reason
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Code Info
GTIN: 00840682105378
Distribution
Worldwide distribution - US Nationwide and the country of China.
Quantity
N/A