FDA Enforcement Class III Terminated

Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

Recall: Z-2133-2014 · Reported August 6, 2014

Enforcement

Recall Number
Z-2133-2014
Event ID
68400
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 6, 2014
Initiation Date
May 27, 2014
Classification Date
July 30, 2014
Termination Date
December 15, 2014
Address
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States

Description

Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

Reason

There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

Code Info

DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX

Distribution

FL LA AR IA OH OH WI IL LA KS MT

Quantity

15