FDA Enforcement
Class III
Terminated
Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
Recall: Z-2133-2014
·
Reported August 6, 2014
Enforcement
- Recall Number
- Z-2133-2014
- Event ID
- 68400
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 6, 2014
- Initiation Date
- May 27, 2014
- Classification Date
- July 30, 2014
- Termination Date
- December 15, 2014
- Address
- 500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States
Description
Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
Reason
There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.
Code Info
DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX
Distribution
FL LA AR IA OH OH WI IL LA KS MT
Quantity
15