I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood.
Recall: Z-2133-2012
·
Reported August 8, 2012
Enforcement
- Recall Number
- Z-2133-2012
- Event ID
- 62371
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Point Of Care Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 8, 2012
- Initiation Date
- January 1, 2011
- Classification Date
- August 2, 2012
- Termination Date
- May 22, 2013
- Address
- 400 College Rd E, N/A, Princeton, NJ, 08540-6607, United States
Description
I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood.
Reason
i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.
Code Info
List number 03P76-25 Lot number S10246A
Distribution
USA ( nationwide )
Quantity
52,750 cartridges