FDA Enforcement Class II Ongoing

Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-36-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 36mm, Novation GXL Liner, 10 Deg Face, 36mm. (note: inadvertently left out of original posting) Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-40-XX, Novation GXL Liner, 10 Deg Face, 20mm ID. (note: inadvertently left out of original posting) Orthopedic hip implant component.

Recall: Z-2132-2021 · Reported July 28, 2021

Enforcement

Recall Number
Z-2132-2021
Event ID
88126
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Exactech, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 28, 2021
Initiation Date
June 29, 2021
Classification Date
July 22, 2021
Address
2320 Nw 66th Ct, N/A, Gainesville, FL, 32653-1630, United States

Description

Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-36-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 36mm, Novation GXL Liner, 10 Deg Face, 36mm. (note: inadvertently left out of original posting) Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-40-XX, Novation GXL Liner, 10 Deg Face, 20mm ID. (note: inadvertently left out of original posting) Orthopedic hip implant component.

Reason

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Code Info

All serial numbers. UDI-DI numbers: 138-32-XX: 10885862024732, 10885862024749, 10885862024756, 10885862024763, 10885862024770. 138-36-XX: 10885862024817, 10885862024824, 10885862024831, 10885862024848, 10885862024886, 10885862024893, 10885862024909, 10885862024916. 138-40-XX: 10885862208804, 10885862208811, 10885862208828.

Distribution

Distribution US nationwide.

Quantity

89,050 total