FDA Enforcement Class II Terminated

MTP Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-B-01001); b) Medium (002-B-02002); c) Large (002-B-03003)

Recall: Z-2131-2018 · Reported June 13, 2018

Enforcement

Recall Number
Z-2131-2018
Event ID
79819
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Orthofix, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 13, 2018
Initiation Date
April 6, 2018
Classification Date
June 7, 2018
Termination Date
June 30, 2021
Address
3451 Plano Pkwy, N/A, Lewisville, TX, 75056-9453, United States

Description

MTP Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-B-01001); b) Medium (002-B-02002); c) Large (002-B-03003)

Reason

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Code Info

(UDI)/Lot: a) (18056099647557) B1098799, B1115575; (18056099647564) B1113628; (18056099647571) B1098800

Distribution

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Quantity

N/A