FDA Enforcement Class II Terminated

94" (239cm) 20 Drop Admin Set w/High Pressure 4-Way (300 psi) Stopcock, Rotating Luer, Non-DEHP Tubing

Recall: Z-2130-2012 · Reported August 8, 2012

Enforcement

Recall Number
Z-2130-2012
Event ID
62538
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 8, 2012
Initiation Date
April 17, 2012
Classification Date
August 2, 2012
Termination Date
August 23, 2012
Address
951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States

Description

94" (239cm) 20 Drop Admin Set w/High Pressure 4-Way (300 psi) Stopcock, Rotating Luer, Non-DEHP Tubing

Reason

The firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded.

Code Info

Part Number B30059 510(k) K964435 Lot number 2418897, 2423996, 2425137, 2430044, 2431861, 2436569, 2443559, 2447238, 2448099, 2452377, 2457453, 2459025, 2464569, 2467687.

Distribution

Nationwide distribution

Quantity

N/A