Porta Shield, Product Number Porta Shield, Part Numbers PTS2, PTS3, PTS4 Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.
Enforcement
- Recall Number
- Z-2129-2019
- Event ID
- 83144
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Burlington Medical, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 7, 2019
- Initiation Date
- June 7, 2019
- Classification Date
- July 30, 2019
- Termination Date
- April 28, 2023
- Address
- PO Box 3194, 3 Elmhurst St, Newport News, VA, 23603-1137, United States
Description
Porta Shield, Product Number Porta Shield, Part Numbers PTS2, PTS3, PTS4 Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.
Serial Numbers: 419886 443251 426551 443014 444704 416997 391132 419882 419877
US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.