FDA Enforcement Class II Terminated

Cardiac Catheterization Waste Management System w/1000mL Waste Bag, 2 Fluid Admin Sets Product Usage: For waste managment.

Recall: Z-2129-2012 · Reported August 8, 2012

Enforcement

Recall Number
Z-2129-2012
Event ID
62538
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 8, 2012
Initiation Date
April 17, 2012
Classification Date
August 2, 2012
Termination Date
August 23, 2012
Address
951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States

Description

Cardiac Catheterization Waste Management System w/1000mL Waste Bag, 2 Fluid Admin Sets Product Usage: For waste managment.

Reason

The firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded.

Code Info

Part Number 460730418 510(k) K052865 Lot number 2444750

Distribution

Nationwide distribution

Quantity

N/A