FDA Enforcement Class II Terminated

High Pressure Stopcock (3-Way, Luer Lock, Off). Product Usage: For IV.

Recall: Z-2128-2012 · Reported August 8, 2012

Enforcement

Recall Number
Z-2128-2012
Event ID
62538
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 8, 2012
Initiation Date
April 17, 2012
Classification Date
August 2, 2012
Termination Date
August 23, 2012
Address
951 Calle Amanecer, N/A, San Clemente, CA, 92673-6212, United States

Description

High Pressure Stopcock (3-Way, Luer Lock, Off). Product Usage: For IV.

Reason

The firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded.

Code Info

Part Number 423530401 510(k) K052865 Lot#s: 2432974, 2441508, 2448504, 2455568

Distribution

Nationwide distribution

Quantity

4625