FDA Enforcement Class II Terminated

Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.

Recall: Z-2127-2020 · Reported June 3, 2020

Enforcement

Recall Number
Z-2127-2020
Event ID
85572
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 3, 2020
Initiation Date
April 17, 2020
Classification Date
May 26, 2020
Termination Date
August 25, 2020
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355, United States

Description

Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.

Reason

The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.

Code Info

Serial Numbers: 11200 13288 13259 12339 13241 10019 12377 13219 12946 13232 13045 12618 12972 40109 40110 40073 40057 40013 40082 40093 40078 40107 40108 40111 40120 40126 40134 40135 40104 40074 40095 40076 40080 40058 40077 40081 40055 40054 40075 40125 40113 40102 40071 40061 40094 40124 40070 40103 40118 40119 40121 40084 40122 40123 40083 40062 40130 40116

Distribution

US Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.

Quantity

58