FDA Enforcement Class II Ongoing

Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-28-XX, Novation GXL Liner, +5mm Lateralized, 28mm; and Novation GXL Liner, +5mm Lateralized, G0, 22mm ID. Orthopedic hip implant component.

Recall: Z-2126-2021 · Reported July 28, 2021

Enforcement

Recall Number
Z-2126-2021
Event ID
88126
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Exactech, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 28, 2021
Initiation Date
June 29, 2021
Classification Date
July 22, 2021
Address
2320 Nw 66th Ct, N/A, Gainesville, FL, 32653-1630, United States

Description

Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-28-XX, Novation GXL Liner, +5mm Lateralized, 28mm; and Novation GXL Liner, +5mm Lateralized, G0, 22mm ID. Orthopedic hip implant component.

Reason

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Code Info

All serial numbers. UDI-DI numbers: 10885862024213, 10885862024220, 10885862024237

Distribution

Distribution US nationwide.

Quantity

89,050 total