Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) LVAD DRIVELINE TRAY, Model Number: DM1035A; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
Enforcement
- Recall Number
- Z-2125-2026
- Event ID
- 98642
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 13, 2026
- Initiation Date
- January 7, 2026
- Classification Date
- May 7, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) LVAD DRIVELINE TRAY, Model Number: DM1035A; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
1) DM1035, UDI-DI: 10653160327752(each), 653160327755(case), Lot Number: 2020042290; 2) DM1035, UDI-DI: 10653160327752(each), 653160327755(case), Lot Number: 2020052290; 3) DM1035, UDI-DI: 10653160327752(each), 653160327755(case), Lot Number: 2020072290; 4) DM1035, UDI-DI: 10653160327752(each), 653160327755(case), Lot Number: 2020082690; 5) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2022030390; 6) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2022072590; 7) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2023062390; 8) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2023101290; 9) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2025031190; 10) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2025051290; 11) DT8670A, UDI-DI: 653160177473(each), 10653160177470(case), Lot Number: 2020051390
Worldwide - US Nationwide distribution.
9720 units