FDA Enforcement Class II Terminated

Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON¿ Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine

Recall: Z-2124-2017 · Reported May 31, 2017

Enforcement

Recall Number
Z-2124-2017
Event ID
77138
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 31, 2017
Initiation Date
April 24, 2017
Classification Date
May 22, 2017
Termination Date
December 30, 2017
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232, United States

Description

Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON¿ Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine

Reason

Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.

Code Info

M609458

Distribution

Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, France, Germany, Guatemala, India, Italy, Japan, Mayotte, Mexico, Netherlands, Oman, Philippines, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.

Quantity

3,158 units total (1,957 units in US)