FDA Enforcement
Class II
Ongoing
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40mm. Orthopedic hip implant component.
Recall: Z-2123-2021
·
Reported July 28, 2021
Enforcement
- Recall Number
- Z-2123-2021
- Event ID
- 88126
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Exactech, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 28, 2021
- Initiation Date
- June 29, 2021
- Classification Date
- July 22, 2021
- Address
- 2320 Nw 66th Ct, N/A, Gainesville, FL, 32653-1630, United States
Description
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40mm. Orthopedic hip implant component.
Reason
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Code Info
All serial numbers. UDI-DI numbers: 10885862207043, 10885862207050, 10885862207067, 10885862303349, 10885862303462, 10885862303363
Distribution
Distribution US nationwide.
Quantity
89,050 total