FDA Enforcement
Class II
Terminated
Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Intended to be used to acquire, store, print, transfer, and archive clinical information.
Recall: Z-2123-2017
·
Reported May 31, 2017
Enforcement
- Recall Number
- Z-2123-2017
- Event ID
- 77124
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Merge Healthcare, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 31, 2017
- Initiation Date
- April 13, 2016
- Classification Date
- May 22, 2017
- Termination Date
- July 17, 2017
- Address
- 900 Walnut Ridge Dr, Hartland, WI, 53029-8347, United States
Description
Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Intended to be used to acquire, store, print, transfer, and archive clinical information.
Reason
Fetal patient report was automatically pulling prior measurement data for a prior fetus since the fetal study is based on the mother's MRN, resulting in the wrong fetal measurements getting referenced.
Code Info
Versions 10.0, 10.0.1, 10.1, 10.1.1,10.1.1.1, and 10.1.2.
Distribution
Distributed to the states of AZ, CT, FL, GA, IL, IN, LA, MI, NY, NC, OH, OK, TX, and VT.
Quantity
30 sites have the potentially affected software versions