FDA Enforcement Class II Ongoing

Medline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A

Recall: Z-2121-2026 · Reported May 13, 2026

Enforcement

Recall Number
Z-2121-2026
Event ID
98642
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 13, 2026
Initiation Date
January 7, 2026
Classification Date
May 7, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A

Reason

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Code Info

1) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 22KBD675; 2) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 23KBA331; 3) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 24ABH890; 4) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 24CBK339; 5) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 24FBK537; 6) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25BBB895; 7) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25BBS071; 8) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25CBB549; 9) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25DBE771; 10) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25EBE995; 11) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25EBG142; 12) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25EBN454; 13) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25FBD161; 14) DYKMBNDL200A, UDI-DI: 10195327240592(each), 40195327240593(case), Lot Number: 25HBD027

Distribution

Worldwide - US Nationwide distribution.

Quantity

108 units