FDA Enforcement
Class III
Terminated
Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.
Recall: Z-2121-2017
·
Reported May 31, 2017
Enforcement
- Recall Number
- Z-2121-2017
- Event ID
- 77083
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medtest Holdings, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 31, 2017
- Initiation Date
- May 15, 2014
- Classification Date
- May 20, 2017
- Termination Date
- June 2, 2017
- Address
- 5449 Research Dr, Canton, MI, 48188-2261, United States
Description
Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.
Reason
The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.
Code Info
Model Number(s): C7590-50 (Chemistry Control Level I), C7591-50 (Chemistry Control Level II), and C7592-100 (Chemistry Control Level I and II).
Distribution
Nationwide
Quantity
132919 Vol vials