Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.

Recall: Z-2118-2014 · Reported August 6, 2014

Enforcement

Recall Number
Z-2118-2014
Event ID
68615
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic Neuromodulation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 6, 2014
Initiation Date
June 11, 2014
Classification Date
July 25, 2014
Termination Date
January 22, 2015
Address
7000 Central Ave Ne, Minneapolis, MN, 55432-3568, United States

Description

Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.

Reason

Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.

Code Info

Lot numbers: 082605614, 082601414, 082602014, 082634313, 082630913, 082629713, 082629613, 082626913, 082623113, 082621413, and 082618213. EXPANDED 10/02/2014 Lot Numbers: 082616214, 082616314, 082618314, and 082618414.

Distribution

Worldwide Distribution - USA including CA, CO, FL, IL, KS, KY, LA, MN, MT, NC, NE, OR, TX, VA, and UT, and Internationally to Canada, Spain, and United Kingdom.

Quantity

300 EXPANDED 10/02/2014 130 devices