FDA Enforcement
Class II
Ongoing
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral, 32mm. Orthopedic hip implant component.
Recall: Z-2117-2021
·
Reported July 28, 2021
Enforcement
- Recall Number
- Z-2117-2021
- Event ID
- 88126
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Exactech, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 28, 2021
- Initiation Date
- June 29, 2021
- Classification Date
- July 22, 2021
- Address
- 2320 Nw 66th Ct, N/A, Gainesville, FL, 32653-1630, United States
Description
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-32-XX, Acumatch GXL 0 Degree Liner, 32mm; and Novation GXL Liner, Neutral, 32mm. Orthopedic hip implant component.
Reason
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Code Info
All serial numbers. UDI-DI numbers: 10885862022097, 10885862022103, 10885862022110, 10885862022127, 10885862207074, 10885862022165, 10885862022172, 10885862022189, 10885862022196
Distribution
Distribution US nationwide.
Quantity
89,050 total