FDA Enforcement Class II Ongoing

Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number:PAIN0150C

Recall: Z-2116-2026 · Reported May 13, 2026

Enforcement

Recall Number
Z-2116-2026
Event ID
98642
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 13, 2026
Initiation Date
January 7, 2026
Classification Date
May 7, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number:PAIN0150C

Reason

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Code Info

1) PAIN0150C, UDI-DI: 10193489819571(each), 40193489819572(case), Lot Number: 21JBV886

Distribution

Worldwide - US Nationwide distribution.

Quantity

448 units