FDA Enforcement
Class III
Terminated
Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.
Recall: Z-2116-2017
·
Reported May 24, 2017
Enforcement
- Recall Number
- Z-2116-2017
- Event ID
- 76985
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medtest Holdings, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- May 24, 2017
- Initiation Date
- February 15, 2012
- Classification Date
- May 18, 2017
- Termination Date
- June 2, 2017
- Address
- 5449 Research Dr, Canton, MI, 48188-2261, United States
Description
Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.
Reason
A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not consistent with the 510k submission package insert.
Code Info
Catalog numbers: HA461, HA761 , HA961 and lot codes:022204, 034801, 102401, 119503, and 134101.
Distribution
US Nationwide Distribution
Quantity
119.168L across All lots