FDA Enforcement Class II Terminated

Accelerator a3600 Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module (P/N FLX-274), Aliquoter Module (P/N FLX-209), Rack Output Module ROM400 (P/N FLX-289), XN-9000 Interface Module (P/N FLX-290), Advia 2120 LAS Interface Module (P/N FLX-219), Vertical Transportation Module (P/N FLX -204), Rack Input Module (P/N FLX-214), AUWi System Interface Module (P/N FLX-286), Track To Rack Interface Module (P/N FLX-295) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Recall: Z-2115-2020 · Reported June 3, 2020

Enforcement

Recall Number
Z-2115-2020
Event ID
85537
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Inpeco S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 3, 2020
Initiation Date
April 7, 2020
Classification Date
May 26, 2020
Termination Date
June 13, 2024
Address
Via San Gottardo 10, N/A, Lugano, N/A, N/A, Switzerland

Description

Accelerator a3600 Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module (P/N FLX-274), Aliquoter Module (P/N FLX-209), Rack Output Module ROM400 (P/N FLX-289), XN-9000 Interface Module (P/N FLX-290), Advia 2120 LAS Interface Module (P/N FLX-219), Vertical Transportation Module (P/N FLX -204), Rack Input Module (P/N FLX-214), AUWi System Interface Module (P/N FLX-286), Track To Rack Interface Module (P/N FLX-295) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Reason

The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.

Code Info

Serial Numbers: ACP.050 ACP.146

Distribution

The products were distributed to the following US States: IL and NY.

Quantity

2 systems