FDA Enforcement
Class II
Terminated
Normed General Instruments 503004541: Werber Countersink Cannulated for micro CBS Screws, AO 503004341: Werber Countersink Cannulated for micro CBS Screws, round-shaft The Countersink is an instrument which is used over the guide wire to prepare for adequate space in the cortical bone rim to sink the screw head in to bones/tissues for different Foot, Ankle and Hand implant systems. The instrument is used in several Normed implant systems. The instrument is used to avoid soft tissue irritation
Recall: Z-2111-2017
·
Reported May 24, 2017
Enforcement
- Recall Number
- Z-2111-2017
- Event ID
- 77262
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Gmbh
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 24, 2017
- Initiation Date
- May 11, 2017
- Classification Date
- May 18, 2017
- Termination Date
- November 3, 2017
- Address
- Sulzer Allee 8, Winterthur, N/A, Switzerland
Description
Normed General Instruments 503004541: Werber Countersink Cannulated for micro CBS Screws, AO 503004341: Werber Countersink Cannulated for micro CBS Screws, round-shaft The Countersink is an instrument which is used over the guide wire to prepare for adequate space in the cortical bone rim to sink the screw head in to bones/tissues for different Foot, Ankle and Hand implant systems. The instrument is used in several Normed implant systems. The instrument is used to avoid soft tissue irritation
Reason
the devices used a different raw material to produce two countersink products which has higher hardness and higher carbon content resulting in device damage.
Code Info
Lots: 13065 13569 13571
Distribution
Single US consignee CO.
Quantity
138 (OUS 297)