FDA Enforcement Class I Terminated

Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.

Recall: Z-2110-2021 · Reported August 18, 2021

Enforcement

Recall Number
Z-2110-2021
Event ID
88073
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 18, 2021
Initiation Date
June 4, 2021
Classification Date
August 6, 2021
Termination Date
September 25, 2024
Address
1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States

Description

Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.

Reason

System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.

Code Info

All Serial Numbers

Distribution

United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.

Quantity

130135 units