FDA Enforcement
Class I
Terminated
Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
Recall: Z-2110-2021
·
Reported August 18, 2021
Enforcement
- Recall Number
- Z-2110-2021
- Event ID
- 88073
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Baxter Healthcare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 18, 2021
- Initiation Date
- June 4, 2021
- Classification Date
- August 6, 2021
- Termination Date
- September 25, 2024
- Address
- 1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States
Description
Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
Reason
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Code Info
All Serial Numbers
Distribution
United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.
Quantity
130135 units