FDA Enforcement Class II Ongoing

See full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 4) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 5) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 6) ROBOTIC-LF, Model Number: CDS984543D; 7) ROBOTIC-LF, Model Number: CDS984543F; 8) ROBOTIC-LF, Model Number: CDS984543G; 9) ROBOTIC-LF, Model Number: CDS984543I; 10) GYN CDS, Model Number: CDS984860F; 11) GYN LAP PELVI, Model Number: CDS985423G; 12) GYN KIT, Model Number: DYKM1657A; 13) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL100; 14) KIT ROBOTICS GYN, Model Number: DYKMBNDL109; 15) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116D; 16) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116F; 17) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119; 18) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119A; 19) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168; 20) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168A; 21) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184D; 22) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91;

Recall: Z-2109-2026 · Reported May 13, 2026

Enforcement

Recall Number
Z-2109-2026
Event ID
98642
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 13, 2026
Initiation Date
January 7, 2026
Classification Date
May 7, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

See full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 4) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 5) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 6) ROBOTIC-LF, Model Number: CDS984543D; 7) ROBOTIC-LF, Model Number: CDS984543F; 8) ROBOTIC-LF, Model Number: CDS984543G; 9) ROBOTIC-LF, Model Number: CDS984543I; 10) GYN CDS, Model Number: CDS984860F; 11) GYN LAP PELVI, Model Number: CDS985423G; 12) GYN KIT, Model Number: DYKM1657A; 13) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL100; 14) KIT ROBOTICS GYN, Model Number: DYKMBNDL109; 15) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116D; 16) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116F; 17) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119; 18) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119A; 19) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168; 20) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168A; 21) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184D; 22) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91;

Reason

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Code Info

1) CDS981504G, UDI-DI: 10195327157968(each), 40195327157969(case), Lot Number: 25BDB744; 2) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23EBE997; 3) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23GBL921; 4) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23GBO574; 5) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23IBC607; 6) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23JBQ095; 7) CDS982893I, UDI-DI: 10889942208050(each), 40889942208051(case), Lot Number: 24IBO240; 8) CDS983300J, UDI-DI: 10193489425048(each), 40193489425049(case), Lot Number: 21CDB781; 9) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21EBA117; 10) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21GBI171; 11) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21HBN807; 12) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21IBM804; 13) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21VBA272; 14) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22BBL450; 15) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22FBV462; 16) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22GBP114; 17) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22HBV491; 18) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22JBC547; 19) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22KBD152; 20) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22OBM422; 21) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23ABC549; 22) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23CBN644; 23) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23CBX440; 24) CDS984543D, UDI-DI: 10195327072322(each), 40195327072323(case), Lot Number: 22CBC587; 25) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22GBD616; 26) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22HMB259; 27) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22HMG671; 28) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22JMA223; 29) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22KMD619; 30) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22OBL925; 31) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 24AMJ461; 32) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23BMD680; 33) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23CMC582; 34) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23DMA508; 35) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23EMI460; 36) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23FMI171; 37) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23GMF425; 38) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23GMH282; 39) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23IMA629; 40) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23JMC376; 41) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23LMA864; 42) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 24BMH483; 43) CDS984543I, UDI-DI: 10195327637071(each), 40195327637072(case), Lot Number: 24CMH050; 44) CDS984860F, UDI-DI: 10193489261974(each), 40193489261975(case), Lot Number: 21ADB205; 45) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21CBH282; 46) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21CBH547; 47) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21DBV243; 48) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25DBN805; 49) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25FBS108; 50) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25HBS386; 51) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25LBN225; 52) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21ABI962; 53) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21BBG514; 54) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21EBA097; 55) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21GBP406; 56) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21JBD206; 57) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABM086; 58) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABO060; 59) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABW347; 60) DYKMBNDL116D, UDI-DI: 10195327608866(each), 40195327608867(case), Lot Number: 24CBF889; 61) DYKMBNDL116D, UDI-DI: 10195327608866(each), 40195327608867(case), Lot Number: 24DBA349; 62) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24DBS388; 63) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24FBC411; 64) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24GBK948; 65) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24IBV184; 66) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24JBB328; 67) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21ABJ213; 68) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21ABL615; 69) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21BBA024; 70) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21BBP181; 71) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21EBG218; 72) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21FBP762; 73) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21HBK337; 74) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21JBC127; 75) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21JBX803; 76) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21LBP646; 77) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21LBY301; 78) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22BBR618; 79) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22CBX517; 80) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22FBI695; 81) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22HBL832; 82) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 22JBC546; 83) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 23KBA328; 84) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 23LBE369; 85) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24ABX678; 86) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24CBJ465; 87) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24DBV608; 88) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24JBD270; 89) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24KBF533; 90) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24KBI207; 91) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25FBR662; 92) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25GBX843; 93) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25HBB957; 94) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21ABI984; 95) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21BBG518; 96) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21EBK837; 97) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21GBP369; 98) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21GBU222; 99) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21IBL095; 100) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21JBJ225; 101) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21KBQ606; 102) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22ABL283; 103) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22ABL299; 104) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22CBX518; 105) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22HBP253; 106) DYKMBNDL168A, UDI-DI: 10195327240806(each), 40195327240807(case), Lot Number: 23FBJ346; 107) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24GBJ568; 108) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24GBV008; 109) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24HBK278; 110) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25ABS671; 111) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25BBF014; 112) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25FBJ143; 113) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25HBC798; 114) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25LBG179; 115) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25LBQ333; 116) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21ABJ086; 117) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21BBG520; 118) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21FBP714; 119) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21HBK341; 120) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21IBO770; 121) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21LBJ751; 122) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 22GBZ017; 123) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 22HBB985; 124) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ABJ861; 125) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21BLA496; 126) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21DLA017; 127) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA256; 128) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA352; 129) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA658; 130) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21HLA598; 131) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21JLA896; 132) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21LLA163; 133) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21LLA284; 134) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ALA125; 135) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ALA331; 136) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22BLA918; 137) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22CLA119; 138) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22DLB036; 139) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ELA227; 140) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ELB163; 141) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22HLA664; 142) DYKMBNDL92A, UDI-DI: 10195327241209(each), 40195327241200(case), Lot Number: 22JLA665; 143) DYKMBNDL92A, UDI-DI: 10195327241209(each), 40195327241200(case), Lot Number: 22KLA731; 144) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24HLB142; 145) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24HLB219; 146) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24JLA881; 147) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24JLB140; 148) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLA194; 149) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLA506; 150) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLB092; 151) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25BLA797; 152) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25DLB036; 153) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25ELA660; 154) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25GLA701; 155) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25HLA275; 156) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25ILA134; 157) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 20LBY985; 158) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21ABM148; 159) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21CBB287; 160) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21CBS045; 161) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21GBB013; 162) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21HBO271; 163) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21IBV704; 164) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 22FBK394; 165) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 22HBH516; 166) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 22LBS837; 167) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABJ591; 168) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABK219; 169) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABO813; 170) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23BBI491; 171) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24DBR903; 172) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24EBV430; 173) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24IBE086; 174) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24KBA746; 175) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24LBN994; 176) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25ABH279; 177) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25ABN371; 178) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25BBA805; 179) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25CBR621; 180) DYKMBNDL94D, UDI-DI: 10198459350641(each), 40198459350642(case), Lot Number: 25EBN175; 181) DYKMBNDL94F, UDI-DI: 10198459468926(each), 40198459468927(case), Lot Number: 25HBT315; 182) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24ABI719; 183) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24BBH633; 184) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24CBM999; 185) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24DBL074; 186) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24EBM495; 187) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24EBT021; 188) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24HBA113; 189) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24HBX130; 190) DYNJ0347372F, UDI-DI: 10195327213176(each), 40195327213177(case), Lot Number: 22GBR392; 191) DYNJ0347372G, UDI-DI: 10195327266028(each), 40195327266029(case), Lot Number: 22KBU586; 192) DYNJ0347372G, UDI-DI: 10195327266028(each), 40195327266029(case), Lot Number: 23DBP818; 193) DYNJ0347372I, UDI-DI: 10195327418335(each), 40195327418336(case), Lot Number: 23GBO695; 194) DYNJ0347372I, UDI-DI: 10195327418335(each), 40195327418336(case), Lot Number: 23IBK088; 195) DYNJ0384702T, UDI-DI: 10195327049683(each), 40195327049684(case), Lot Number: 23BBN374; 196) DYNJ28252G, UDI-DI: 10198459088124(each), 40198459088125(case), Lot Number: 25BBP572; 197) DYNJ28252G, UDI-DI: 10198459088124(each), 40198459088125(case), Lot Number: 25CBC137; 198) DYNJ30005G, UDI-DI: 10193489601794(each), 40193489601795(case), Lot Number: 21DMA051; 199) DYNJ32706B, UDI-DI: 10195327364304(each), 40195327364305(case), Lot Number: 23EMB628; 200) DYNJ38844N, UDI-DI: 10195327243166(each), 40195327243167(case), Lot Number: 23DBI585; 201) DYNJ38844N, UDI-DI: 10195327243166(each), 40195327243167(case), Lot Number: 23HBI619; 202) DYNJ47550C, UDI-DI: 10195327209964(each), 40195327209965(case), Lot Number: 23DBD093; 203) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23BBC881; 204) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23CBR808; 205) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23FBH830; 206) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23FBU806; 207) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23HBO785; 208) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23IBQ207; 209) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23JBN605; 210) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23KBQ430; 211) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24CBQ847; 212) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24FBG329; 213) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24HBV800; 214) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24IBE460; 215) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24JBC084; 216) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24KBM397; 217) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25CBA290; 218) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25DBE603; 219) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25DBE964; 220) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25KBF492; 221) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21AMA809; 222) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21GMD682; 223) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21JMF391; 224) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21JMH124; 225) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21JMH124A; 226) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21KMG274; 227) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22BMH522; 228) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22EMA211; 229) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22HMA973; 230) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22HMC351; 231) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22KMH193; 232) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 23CMB495; 233) DYNJ49705K, UDI-DI: 10195327020347(each), 40195327020348(case), Lot Number: 22IMI306; 234) DYNJ50348F, UDI-DI: 10195327406349(each), 40195327406340(case), Lot Number: 23IMD230; 235) DYNJ50439F, UDI-DI: 10193489704921(each), 40193489704922(case), Lot Number: 22EBC737; 236) DYNJ51795F, UDI-DI: 10193489475937(each), 40193489475938(case), Lot Number: 20LDC148; 237) DYNJ51795F, UDI-DI: 10193489475937(each), 40193489475938(case), Lot Number: 21ADA184; 238) DYNJ52569B, UDI-DI: 10193489685190(each), 40193489685191(case), Lot Number: 23EBS784; 239) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21CBU815; 240) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21IBG462; 241) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21JBR818; 242) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22ABC476; 243) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22GBJ776; 244) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22HBV228; 245) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22LBD784; 246) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 23BBA462; 247) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21EMB589; 248) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21FMB889; 249) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21GME150; 250) DYNJ58321A, UDI-DI: 10193489489316(each), 40193489489317(case), Lot Number: 22ABG197; 251) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21DMB839; 252) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21EMF383; 253) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21HMD002; 254) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21JMF224; 255) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 22AME373; 256) DYNJ63079C, UDI-DI: 10195327102180(each), 40195327102181(case), Lot Number: 23EMI624; 257) DYNJ63079C, UDI-DI: 10195327102180(each), 40195327102181(case), Lot Number: 23GMB716; 258) DYNJ63079D, UDI-DI: 10195327505677(each), 40195327505678(case), Lot Number: 23JMB435; 259) DYNJ63079D, UDI-DI: 10195327505677(each), 40195327505678(case), Lot Number: 24BME368; 260) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24HMC603; 261) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24HMH481; 262) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24JMD681; 263) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25AMH256; 264) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25JMH582; 265) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25JMK630; 266) DYNJ64266, UDI-DI: 10193489795516(each), 40193489795517(case), Lot Number: 21ADA447; 267) DYNJ66298D, UDI-DI: 10195327095895(each), 40195327095896(case), Lot Number: 23IBW233; 268) DYNJ67269, UDI-DI: 10193489394696(each), 40193489394697(case), Lot Number: 22KBG504; 269) DYNJ68710, UDI-DI: 10193489842906(each), 40193489842907(case), Lot Number: 22ABM641; 270) DYNJ69419A, UDI-DI: 10195327223588(each), 40195327223589(case), Lot Number: 23ABS144; 271) DYNJ69419A, UDI-DI: 10195327223588(each), 40195327223589(case), Lot Number: 23DBK434; 272) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24EMC427; 273) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24EMD677; 274) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24GMC135; 275) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24GMG805; 276) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24KME386; 277) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24KMH362; 278) DYNJ81492A, UDI-DI: 10195327348830(each), 40195327348831(case), Lot Number: 23BBT116; 279) DYNJ81761A, UDI-DI: 10195327574468(each), 40195327574469(case), Lot Number: 24FMF687; 280) DYNJ83475A, UDI-DI: 10195327460938(each), 40195327460939(case), Lot Number: 23HBY488; 281) DYNJ84783, UDI-DI: 10195327417949(each), 40195327417940(case), Lot Number: 23IBU517; 282) DYNJ85342, UDI-DI: 10195327491437(each), 40195327491438(case), Lot Number: 23KMF007; 283) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 24EBD237; 284) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 25ABP819; 285) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 25ABT160; 286) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 25ABT161; 287) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 25GBA444; 288) DYNJ900262A, UDI-DI: 10195327477820(each), 40195327477821(case), Lot Number: 23KBU736; 289) DYNJ901046L, UDI-DI: 10195327458614(each), 40195327458615(case), Lot Number: 23KBF626; 290) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21JBM245; 291) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21JBV891; 292) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21KBF079; 293) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21KBP568; 294) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 22JBR884; 295) DYNJ901559J, UDI-DI: 10193489235890(each), 40193489235891(case), Lot Number: 21CBH289; 296) DYNJ901559J, UDI-DI: 10193489235890(each), 40193489235891(case), Lot Number: 21EBR740; 297) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21ADB190; 298) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21BDA744; 299) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21FDB236; 300) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25ABS755; 301) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25DBB182; 302) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25EBK653; 303) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25GBR954; 304) DYNJ905066B, UDI-DI: 10195327098674(each), 40195327098675(case), Lot Number: 23BMB931; 305) DYNJ905066B, UDI-DI: 10195327098674(each), 40195327098675(case), Lot Number: 23BMF722; 306) DYNJ905269G, UDI-DI: 10193489337112(each), 40193489337113(case), Lot Number: 21ABH751; 307) DYNJ905269G, UDI-DI: 10193489337112(each), 40193489337113(case), Lot Number: 21BBD394; 308) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 22GBL545; 309) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23DBT141; 310) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23EBQ671; 311) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23FBE986; 312) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23BBA689; 313) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23BBB047; 314) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23DBM357; 315) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23FBP128; 316) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23HBO123; 317) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23IBO509; 318) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23JBT789; 319) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23LBJ184; 320) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24ABD306; 321) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24BBN302; 322) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24CBS248; 323) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24DBT181; 324) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24FBJ836; 325) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24HBM625; 326) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24IBN246; 327) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24KBJ970; 328) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24LBN196; 329) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25ABT788; 330) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25BBO090; 331) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25EBA889; 332) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25GBB683; 333) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25HBO792; 334) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25IBI139; 335) DYNJ906452F, UDI-DI: 10198459575303(each), 40198459575304(case), Lot Number: 25KBG657; 336) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23BBU275; 337) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23DBM354; 338) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23EBN137; 339) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23GBH151; 340) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23IBH652; 341) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23JBB062; 342) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23KBW414; 343) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23LBN387; 344) DYNJ907753F, UDI-DI: 10195327194482(each), 40195327194483(case), Lot Number: 22LBN185; 345) DYNJ907771C, UDI-DI: 10195327120399(each), 40195327120390(case), Lot Number: 22LBP170; 346) DYNJ908887B, UDI-DI: 10195327620158(each), 40195327620159(case), Lot Number: 24CBP744; 347) DYNJ908887B, UDI-DI: 10195327620158(each), 40195327620159(case), Lot Number: 24DBD796; 348) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24EBM187; 349) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24GBK492; 350) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24HBI648; 351) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24JBP902; 352) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24LBP187; 353) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23HMI269; 354) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23JMB528; 355) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23KME175; 356) DYNJ909799, UDI-DI: 10195327483210(each), 40195327483211(case), Lot Number: 23LBC746; 357) DYNJ909799, UDI-DI: 10195327483210(each), 40195327483211(case), Lot Number: 24HBE897; 358) DYNJ910009,

More Code Info

UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24BMA251; 359) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24CMI706; 360) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24DMI073; 361) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24HMA202; 362) DYNJ910009A, UDI-DI: 10198459122507(each), 40198459122508(case), Lot Number: 24KMD940; 363) DYNJ910009B, UDI-DI: 10198459240133(each), 40198459240134(case), Lot Number: 25FMD839; 364) DYNJ910476, UDI-DI: 10198459009174(each), 40198459009175(case), Lot Number: 24FMJ212; 365) DYNJ910536, UDI-DI: 10198459029554(each), 40198459029555(case), Lot Number: 24GBA620; 366) DYNJ910536A, UDI-DI: 10198459054358(each), 40198459054359(case), Lot Number: 24IBE833; 367) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 24JBS568; 368) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 24LBM436; 369) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25BBD998; 370) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25DBM184; 371) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25EBD405; 372) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25EBL131; 373) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25FBV527; 374) DYNJ910589, UDI-DI: 10198459040894(each), 40198459040895(case), Lot Number: 24HBF839; 375) DYNJ910594, UDI-DI: 10198459041327(each), 40198459041328(case), Lot Number: 24HBF837; 376) DYNJ911567, UDI-DI: 10198459409004(each), 40198459409005(case), Lot Number: 25HBB593; 377) DYNJ911567, UDI-DI: 10198459409004(each), 40198459409005(case), Lot Number: 25HBJ084; 378) DYNJ911568, UDI-DI: 10198459409011(each), 40198459409012(case), Lot Number: 25HBB594; 379) DYNJ9682698S, UDI-DI: 10195327257170(each), 40195327257171(case), Lot Number: 23IDC305; 380) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21CKA647; 381) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21CKA812; 382) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21DMF113; 383) DYNJ9701096I, UDI-DI: 10193489939651(each), 40193489939652(case), Lot Number: 22AMC654; 384) DYNJT3381, UDI-DI: 10198459325205(each), 40198459325206(case), Lot Number: 25EBK724; 385) DYNJT7540, UDI-DI: 10198459630989(each), 40198459630980(case), Lot Number: 26ABD262

Distribution

Worldwide - US Nationwide distribution.

Quantity

15806 units