FDA Enforcement Class II Terminated

Big Bore Oncology tomography X-ray system scanner, model number 728243. The Big Bore Oncology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes

Recall: Z-2109-2016 · Reported July 6, 2016

Enforcement

Recall Number
Z-2109-2016
Event ID
74507
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 6, 2016
Initiation Date
June 14, 2016
Classification Date
June 30, 2016
Termination Date
July 2, 2018
Address
595 Miner Rd, Cleveland, OH, 44143-2131, United States

Description

Big Bore Oncology tomography X-ray system scanner, model number 728243. The Big Bore Oncology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes

Reason

Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.

Code Info

Big Bore Oncology System Code 728243; Serial numbers: 75002, 75003, 75004, 75005, 75006, 75007, 75008, 75009, 75010, 75011, 75012, 75014, 75015, 75016, 75017, 75018, 75019, 75020, 75021, 75022, 75023, 75024, 75025, 75026, 75030, 75032, 75035, 75036, 75037, 75040, 75042, 75044, 75047, 75048, 75049, 75051, 75053, 75060, 75064, 75100, 75102, 75104, 75105, 750111, 750114

Distribution

Worldwide distribution to Belgium, China, Egypt, Germany, India, Korea, Republic of, Myanmar, Nigeria, Peru, Russian Federation, Rwanda, Singapore, Spain, Sweden, Taiwan, Thailand, and United Kingdom

Quantity

47 units