FDA Enforcement Class II Terminated

QLAB 10 Core Module PN 453561704771. QLAB Quantification Software is a software application package designed to view and quantify image data acquired on Philips Medical Systems ultrasound products.

Recall: Z-2109-2014 · Reported July 30, 2014

Enforcement

Recall Number
Z-2109-2014
Event ID
68756
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Ultrasound, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 30, 2014
Initiation Date
July 9, 2014
Classification Date
July 23, 2014
Termination Date
November 6, 2015
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431, United States

Description

QLAB 10 Core Module PN 453561704771. QLAB Quantification Software is a software application package designed to view and quantify image data acquired on Philips Medical Systems ultrasound products.

Reason

The QLAB a2DQ and aCMQ features have a defect that may use incorrect measurement values under specific conditions.

Code Info

*****SYSTEM BASED ID OUTSIDE THE US********** 120F-P213FA35-V41E, 279E-Q21384E8-P41D, 3FBF-X2649F25-J4BF, 47FD-Q1FD112B-J3F0, 4A8F-U24EA0B5-R493, 5570-Q20D1AB2-W410, 5D8C-G1FC9BDE-N3EF, 684B-V2122501-R41A, 6DA2-G232AD5C-S45B, 731B-Q1FC2691-V3EE, 808E-P20C3018-I40E, 88AA-T1FBB144-W3ED, A0CB-I24CCB81-M48F, A0ED-H1F5BC5B-P3E1, AB9B-S236CD11-W463, BE98-V1E4C83A-P3BF, C6A3-H1FFD0F9-X3F5, CBE9-M24BE0E7-Y48D, EEFC-Y22FED8E-S455, EF0D-I20465FB-I3FE, F475-U1F966C3-X3E8, and V49C-R203F0AE-I3FE. ***********Two units shipped in US and one unit shipped outside the US to distributors do not have serial numbers. There are licensing agreements with these distributors that allows the distributors to create their own licensing and serial numbers and/or tracking information.

Distribution

Distributed in the states of CA, IA, KY, MD, NY, PA, UT, VA, WA and WI. Also distributed in Australia, Belgium, Canada, India, Italy, Japan, South Korea, Netherlands, New Zealand, Spain, Sweden, and Switzerland.

Quantity

25 units total (2 units in the US and 23 units outside the US)