FDA Enforcement Class II Ongoing

Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38B; 6) L & D PACK-LF, Model Number: DYNJ0456816O; 7) L & D PACK-LF, Model Number: DYNJ0456816P; 8) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 10) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 11) DELIVERY PACK-LF, Model Number: DYNJ20546F; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 13) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 14) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 15) OB PACK-LF, Model Number: DYNJ26121I; 16) OB DELIVERY PACK, Model Number: DYNJ31024J; 17) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 18) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 19) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 20) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 21) VAG DELIVERY PACK, Model Number: DYNJ38481G; 22) VAG DELIVERY PACK, Model Number: DYNJ38481I; 23) VAG DELIVERY PACK, Model Number: DYNJ38481K; 24) OB PACK 2, Model Number: DYNJ39000D; 25) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 26) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 27) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 28) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 29) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 31) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 32) DELIVERY PACK, Model Number: DYNJ55373A; 33) VAGINAL DELIVERY PACK, Model Number: DYNJ59403F; 34) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 35) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 36) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 37) OB-VAG PACK 9128190, Model Number: DYNJ62341; 38) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 39) LABOR AND DELIVERY TRAY, Model Number: DYNJ62722B; 40) L & D PACK, Model Number: DYNJ63216A; 41) L&D PACK, Model Number: DYNJ63757A; 42) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 43) MINOR VAG PACK, Model Number: DYNJ68663; 44) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 45) GYN VAGINAL PACK, Model Number: DYNJ68681; 46) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 47) VAGINAL PACK, Model Number: DYNJ69774A; 48) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 49) DELIVERY PACK, Model Number: DYNJ87469A; 50) DELIVERY B, Model Number: DYNJ900351T; 51) VAGINAL DELIVERY, Model Number: DYNJ901387A; 52) FLOYD DELIVERY, Model Number: DYNJ906937D; 53) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 54) OB PACK, Model Number: DYNJT5186; 55) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 56) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 57) LABOR AND DELIVERY KIT, Model Number: MNS8950A

Recall: Z-2108-2026 · Reported May 13, 2026

Enforcement

Recall Number
Z-2108-2026
Event ID
98642
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 13, 2026
Initiation Date
January 7, 2026
Classification Date
May 7, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38B; 6) L & D PACK-LF, Model Number: DYNJ0456816O; 7) L & D PACK-LF, Model Number: DYNJ0456816P; 8) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 10) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 11) DELIVERY PACK-LF, Model Number: DYNJ20546F; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 13) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 14) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 15) OB PACK-LF, Model Number: DYNJ26121I; 16) OB DELIVERY PACK, Model Number: DYNJ31024J; 17) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 18) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 19) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 20) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 21) VAG DELIVERY PACK, Model Number: DYNJ38481G; 22) VAG DELIVERY PACK, Model Number: DYNJ38481I; 23) VAG DELIVERY PACK, Model Number: DYNJ38481K; 24) OB PACK 2, Model Number: DYNJ39000D; 25) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 26) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 27) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 28) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 29) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 31) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 32) DELIVERY PACK, Model Number: DYNJ55373A; 33) VAGINAL DELIVERY PACK, Model Number: DYNJ59403F; 34) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 35) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 36) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 37) OB-VAG PACK 9128190, Model Number: DYNJ62341; 38) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 39) LABOR AND DELIVERY TRAY, Model Number: DYNJ62722B; 40) L & D PACK, Model Number: DYNJ63216A; 41) L&D PACK, Model Number: DYNJ63757A; 42) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 43) MINOR VAG PACK, Model Number: DYNJ68663; 44) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 45) GYN VAGINAL PACK, Model Number: DYNJ68681; 46) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 47) VAGINAL PACK, Model Number: DYNJ69774A; 48) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 49) DELIVERY PACK, Model Number: DYNJ87469A; 50) DELIVERY B, Model Number: DYNJ900351T; 51) VAGINAL DELIVERY, Model Number: DYNJ901387A; 52) FLOYD DELIVERY, Model Number: DYNJ906937D; 53) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 54) OB PACK, Model Number: DYNJT5186; 55) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 56) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 57) LABOR AND DELIVERY KIT, Model Number: MNS8950A

Reason

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Code Info

1) DYKM2326, UDI-DI: 10198459565748(each), 40198459565749(case), Lot Number: 25KMF765; 2) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21ABI982; 3) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21ABT654; 4) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21DBA335; 5) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21EBE875; 6) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21FBP760; 7) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21GBN267; 8) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21IBE978; 9) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21IBO498; 10) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21JBQ910; 11) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21KBM483; 12) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 21LBX453; 13) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 22IBB499; 14) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 22KBS631; 15) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 22LBB693; 16) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 23BBT994; 17) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 23EBP170; 18) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 23HBR548; 19) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 23KBA329; 20) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 23KBV905; 21) DYKMBNDL165, UDI-DI: 10193489477405(each), 40193489477406(case), Lot Number: 24ABK386; 22) DYKMBNDL31, UDI-DI: 10193489478716(each), 40193489478717(case), Lot Number: 21ABC201; 23) DYKMBNDL31, UDI-DI: 10193489478716(each), 40193489478717(case), Lot Number: 23JLA066; 24) DYKMBNDL31, UDI-DI: 10193489478716(each), 40193489478717(case), Lot Number: 23JLA538; 25) DYKMBNDL38, UDI-DI: 10193489477757(each), 40193489477758(case), Lot Number: 21ABM087; 26) DYKMBNDL38, UDI-DI: 10193489477757(each), 40193489477758(case), Lot Number: 23IBT256; 27) DYKMBNDL38, UDI-DI: 10193489477757(each), 40193489477758(case), Lot Number: 23JBB705; 28) DYKMBNDL38B, UDI-DI: 10195327660017(each), 40195327660018(case), Lot Number: 24IBK537; 29) DYKMBNDL38B, UDI-DI: 10195327660017(each), 40195327660018(case), Lot Number: 24IBM370; 30) DYKMBNDL38B, UDI-DI: 10195327660017(each), 40195327660018(case), Lot Number: 24IBW265; 31) DYKMBNDL38B, UDI-DI: 10195327660017(each), 40195327660018(case), Lot Number: 24JBK520; 32) DYNJ0456816O, UDI-DI: 10195327435219(each), 40195327435210(case), Lot Number: 23JMC541; 33) DYNJ0456816O, UDI-DI: 10195327435219(each), 40195327435210(case), Lot Number: 23JME434; 34) DYNJ0456816P, UDI-DI: 10195327570545(each), 40195327570546(case), Lot Number: 24HMK119; 35) DYNJ04692S, UDI-DI: 10193489338454(each), 40193489338455(case), Lot Number: 21ADB239; 36) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 23KMJ818; 37) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 24BMA554; 38) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 24CMB999; 39) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 24DMI443; 40) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 24FMF250; 41) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 24GMJ484; 42) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 24IMF157; 43) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 24JMB353; 44) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 25AME014; 45) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 25CME932; 46) DYNJ16621B, UDI-DI: 10195327558116(each), 40195327558117(case), Lot Number: 25DMH878; 47) DYNJ16621C, UDI-DI: 10198459421297(each), 40198459421298(case), Lot Number: 25GMD428; 48) DYNJ16621C, UDI-DI: 10198459421297(each), 40198459421298(case), Lot Number: 25HMC583; 49) DYNJ20546F, UDI-DI: 10193489346084(each), 40193489346085(case), Lot Number: 22IMG976; 50) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 21CME149; 51) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 21DMD921; 52) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 21GMB531; 53) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 21HMD738; 54) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 21IMD851; 55) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 21JMA483; 56) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 21KMB071; 57) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 22AMA608; 58) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 22CME094; 59) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 22DMB332; 60) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 22EMF465; 61) DYNJ20979I, UDI-DI: 10193489865905(each), 40193489865906(case), Lot Number: 22FMB686; 62) DYNJ20979J, UDI-DI: 10195327191733(each), 40195327191734(case), Lot Number: 22GMH899; 63) DYNJ20979J, UDI-DI: 10195327191733(each), 40195327191734(case), Lot Number: 22IMH408; 64) DYNJ20979J, UDI-DI: 10195327191733(each), 40195327191734(case), Lot Number: 22LMA315; 65) DYNJ20979J, UDI-DI: 10195327191733(each), 40195327191734(case), Lot Number: 23AMJ134; 66) DYNJ24108D, UDI-DI: 10195327425258(each), 40195327425259(case), Lot Number: 23GBU622; 67) DYNJ24108D, UDI-DI: 10195327425258(each), 40195327425259(case), Lot Number: 23IBA365; 68) DYNJ24108D, UDI-DI: 10195327425258(each), 40195327425259(case), Lot Number: 23IBS023; 69) DYNJ26121I, UDI-DI: 10198459329517(each), 40198459329518(case), Lot Number: 25EBA350; 70) DYNJ26121I, UDI-DI: 10198459329517(each), 40198459329518(case), Lot Number: 25FBE718; 71) DYNJ26121I, UDI-DI: 10198459329517(each), 40198459329518(case), Lot Number: 25GBJ925; 72) DYNJ26121I, UDI-DI: 10198459329517(each), 40198459329518(case), Lot Number: 25HBM068; 73) DYNJ26121I, UDI-DI: 10198459329517(each), 40198459329518(case), Lot Number: 25IBJ589; 74) DYNJ31024J, UDI-DI: 10195327139544(each), 40195327139545(case), Lot Number: 22IMF645; 75) DYNJ32139C, UDI-DI: 10193489588576(each), 40193489588577(case), Lot Number: 25DBK245; 76) DYNJ33590F, UDI-DI: 10193489220759(each), 40193489220750(case), Lot Number: 20LDC125; 77) DYNJ35488F, UDI-DI: 10193489803488(each), 40193489803489(case), Lot Number: 23JMB591; 78) DYNJ37331C, UDI-DI: 10889942670512(each), 40889942670513(case), Lot Number: 25GBN237; 79) DYNJ38481G, UDI-DI: 10193489727937(each), 40193489727938(case), Lot Number: 20LMD360; 80) DYNJ38481I, UDI-DI: 10193489825251(each), 40193489825252(case), Lot Number: 21DMB477; 81) DYNJ38481I, UDI-DI: 10193489825251(each), 40193489825252(case), Lot Number: 21GMB665; 82) DYNJ38481I, UDI-DI: 10193489825251(each), 40193489825252(case), Lot Number: 21IMB736; 83) DYNJ38481I, UDI-DI: 10193489825251(each), 40193489825252(case), Lot Number: 21JMC579; 84) DYNJ38481I, UDI-DI: 10193489825251(each), 40193489825252(case), Lot Number: 21JME107; 85) DYNJ38481I, UDI-DI: 10193489825251(each), 40193489825252(case), Lot Number: 21LMB210; 86) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 22CMA009; 87) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 22CMB285; 88) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 22EME995; 89) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 22GMH656; 90) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 22IMG983; 91) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23AMC268; 92) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23AMC551; 93) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23AMC555; 94) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23AMJ499; 95) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23CMF409; 96) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23EMD649; 97) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23GMD513; 98) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23HMC301; 99) DYNJ38481K, UDI-DI: 10195327118242(each), 40195327118243(case), Lot Number: 23IMD484; 100) DYNJ39000D, UDI-DI: 10195327319656(each), 40195327319657(case), Lot Number: 23JMB120; 101) DYNJ40203A, UDI-DI: 10193489670998(each), 40193489670999(case), Lot Number: 21GDA744; 102) DYNJ40918A, UDI-DI: 10195327067847(each), 40195327067848(case), Lot Number: 22EBD699; 103) DYNJ40918A, UDI-DI: 10195327067847(each), 40195327067848(case), Lot Number: 22FBP490; 104) DYNJ40918A, UDI-DI: 10195327067847(each), 40195327067848(case), Lot Number: 22HBL591; 105) DYNJ40918A, UDI-DI: 10195327067847(each), 40195327067848(case), Lot Number: 23ABO451; 106) DYNJ40918A, UDI-DI: 10195327067847(each), 40195327067848(case), Lot Number: 23EBQ232; 107) DYNJ40918A, UDI-DI: 10195327067847(each), 40195327067848(case), Lot Number: 23FBM020; 108) DYNJ40918A, UDI-DI: 10195327067847(each), 40195327067848(case), Lot Number: 23GBM044; 109) DYNJ44720B, UDI-DI: 10195327277512(each), 40195327277513(case), Lot Number: 23FDB511; 110) DYNJ51377G, UDI-DI: 10193489906325(each), 40193489906326(case), Lot Number: 23HBO866; 111) DYNJ51794B, UDI-DI: 10193489644708(each), 40193489644709(case), Lot Number: 21ADA465; 112) DYNJ55363A, UDI-DI: 10889942756742(each), 40889942756743(case), Lot Number: 21ABH443; 113) DYNJ55363A, UDI-DI: 10889942756742(each), 40889942756743(case), Lot Number: 21BBB866; 114) DYNJ55363A, UDI-DI: 10889942756742(each), 40889942756743(case), Lot Number: 21JBY447; 115) DYNJ55363A, UDI-DI: 10889942756742(each), 40889942756743(case), Lot Number: 21LBT403; 116) DYNJ55363A, UDI-DI: 10889942756742(each), 40889942756743(case), Lot Number: 22CBW139; 117) DYNJ55363A, UDI-DI: 10889942756742(each), 40889942756743(case), Lot Number: 22CBW142; 118) DYNJ55363A, UDI-DI: 10889942756742(each), 40889942756743(case), Lot Number: 22EBI837; 119) DYNJ55363A, UDI-DI: 10889942756742(each), 40889942756743(case), Lot Number: 22OBB980; 120) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 22FBO672; 121) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 22HBA092; 122) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 22HBM753; 123) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 22HBM755; 124) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 22IBQ631; 125) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 22JBD960; 126) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 23ABO534; 127) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 23CBA634; 128) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 23EBJ019; 129) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 23FBE833; 130) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 23GBN972; 131) DYNJ55363B, UDI-DI: 10195327173548(each), 40195327173549(case), Lot Number: 23HBT039; 132) DYNJ55373A, UDI-DI: 10889942690947(each), 40889942690948(case), Lot Number: 23FBV941; 133) DYNJ59403F, UDI-DI: 10195327441999(each), 40195327441990(case), Lot Number: 23HMG011; 134) DYNJ61269B, UDI-DI: 10195327054687(each), 40195327054688(case), Lot Number: 22OBD693; 135) DYNJ61574, UDI-DI: 10193489601084(each), 40193489601085(case), Lot Number: 23BMF067; 136) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 21CBU863; 137) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 21GBK690; 138) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 21IBA396; 139) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 21KBV669; 140) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 21LBT375; 141) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 21LBT378; 142) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 21VBA541; 143) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 22HBT023; 144) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 22JBW013; 145) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 23DBL820; 146) DYNJ61650, UDI-DI: 10193489586053(each), 40193489586054(case), Lot Number: 23FBR019; 147) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 20LDB518; 148) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 21BDC025; 149) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 21EDB976; 150) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 21FDC626; 151) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 21IDA188; 152) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 22DDC251; 153) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 22EDB025; 154) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 22IDA469; 155) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 23BDA994; 156) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 23DDB664; 157) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 23EDA759; 158) DYNJ62341, UDI-DI: 10193489647020(each), 40193489647021(case), Lot Number: 23FDA173; 159) DYNJ62402A, UDI-DI: 10193489926132(each), 40193489926133(case), Lot Number: 23HBL396; 160) DYNJ62402A, UDI-DI: 10193489926132(each), 40193489926133(case), Lot Number: 23IBS478; 161) DYNJ62402A, UDI-DI: 10193489926132(each), 40193489926133(case), Lot Number: 23KBF906; 162) DYNJ62722B, UDI-DI: 10195327292157(each), 40195327292158(case), Lot Number: 25BBA614; 163) DYNJ63216A, UDI-DI: 10198459065125(each), 40198459065126(case), Lot Number: 24HBN118; 164) DYNJ63216A, UDI-DI: 10198459065125(each), 40198459065126(case), Lot Number: 24KBM262; 165) DYNJ63216A, UDI-DI: 10198459065125(each), 40198459065126(case), Lot Number: 25ABR935; 166) DYNJ63216A, UDI-DI: 10198459065125(each), 40198459065126(case), Lot Number: 25CBL912; 167) DYNJ63216A, UDI-DI: 10198459065125(each), 40198459065126(case), Lot Number: 25GBA096; 168) DYNJ63216A, UDI-DI: 10198459065125(each), 40198459065126(case), Lot Number: 25HBG200; 169) DYNJ63216A, UDI-DI: 10198459065125(each), 40198459065126(case), Lot Number: 25JBU638; 170) DYNJ63757A, UDI-DI: 10195327542481(each), 40195327542482(case), Lot Number: 23KBG966; 171) DYNJ65145, UDI-DI: 10193489247848(each), 40193489247849(case), Lot Number: 21IDB501; 172) DYNJ65145, UDI-DI: 10193489247848(each), 40193489247849(case), Lot Number: 21KDA784; 173) DYNJ65145, UDI-DI: 10193489247848(each), 40193489247849(case), Lot Number: 21LDB251; 174) DYNJ65145, UDI-DI: 10193489247848(each), 40193489247849(case), Lot Number: 22DDA962; 175) DYNJ68663, UDI-DI: 10193489837346(each), 40193489837347(case), Lot Number: 21LBB336; 176) DYNJ68673A, UDI-DI: 10198459221712(each), 40198459221713(case), Lot Number: 25GBC048; 177) DYNJ68681, UDI-DI: 10193489840209(each), 40193489840200(case), Lot Number: 22ABN829; 178) DYNJ69544A, UDI-DI: 10195327325367(each), 40195327325368(case), Lot Number: 25DBK232; 179) DYNJ69774A, UDI-DI: 10195327328368(each), 40195327328369(case), Lot Number: 23IBU502; 180) DYNJ69774A, UDI-DI: 10195327328368(each), 40195327328369(case), Lot Number: 23KBM418; 181) DYNJ84393, UDI-DI: 10195327361990(each), 40195327361991(case), Lot Number: 25HMC554; 182) DYNJ87469A, UDI-DI: 10198459336522(each), 40198459336523(case), Lot Number: 25GBV520; 183) DYNJ900351T, UDI-DI: 10193489818482(each), 40193489818483(case), Lot Number: 22LMI886; 184) DYNJ901387A, UDI-DI: 10193489707922(each), 40193489707923(case), Lot Number: 24IBP827; 185) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23ABR935; 186) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23CBC332; 187) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23CBU513; 188) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23EBU945; 189) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23GBE943; 190) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23GBV967; 191) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23HBQ793; 192) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23IBK055; 193) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23JBX533; 194) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 23LBR077; 195) DYNJ906937D, UDI-DI: 10193489963335(each), 40193489963336(case), Lot Number: 24ABB990; 196) DYNJT3445, UDI-DI: 10198459325175(each), 40198459325176(case), Lot Number: 25EBI391; 197) DYNJT3445, UDI-DI: 10198459325175(each), 40198459325176(case), Lot Number: 25HBO297; 198) DYNJT3445, UDI-DI: 10198459325175(each), 40198459325176(case), Lot Number: 25IBD649; 199) DYNJT3445, UDI-DI: 10198459325175(each), 40198459325176(case), Lot Number: 25IBO981; 200) DYNJT3445, UDI-DI: 10198459325175(each), 40198459325176(case), Lot Number: 25JBR855; 201) DYNJT5186, UDI-DI: 10198459468797(each), 40198459468798(case), Lot Number: 25HBM919; 202) MNS13550A, UDI-DI: 10653160992875(each), 40653160992876(case), Lot Number: 23JBU943; 203) MNS13550A, UDI-DI: 10653160992875(each), 40653160992876(case), Lot Number: 25DBJ182; 204) MNS7460A, UDI-DI: 10653160996392(each), 40653160996393(case), Lot Number: 24CBN067; 205) MNS7460A, UDI-DI: 10653160996392(each), 40653160996393(case), Lot Number: 24GBK561; 206) MNS7460A, UDI-DI: 10653160996392(each), 40653160996393(case), Lot Number: 24HBA805; 207) MNS7460A, UDI-DI: 10653160996392(each), 40653160996393(case), Lot Number: 24IBC061; 208) MNS7460A, UDI-DI: 10653160996392(each), 40653160996393(case), Lot Number: 24JBU227; 209) MNS7460A, UDI-DI: 10653160996392(each), 40653160996393(case), Lot Number: 25ABJ839; 210) MNS7460A, UDI-DI: 10653160996392(each), 40653160996393(case), Lot Number: 25ABQ083; 211) MNS7460A, UDI-DI: 10653160996392(each), 40653160996393(case), Lot Number: 25DBK156; 212) MNS8950A, UDI-DI: 10653160320388(each), 40653160320389(case), Lot Number: 21GBG870

Distribution

Worldwide - US Nationwide distribution.

Quantity

29192