Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table top (1180-16XX) serves to support and position patients for surgical procedures (e.g. cardiovascular surgery) and interventional examination immediately before, during and after the surgical phase. The table top is radiotranslucent and enables intraoperative use of X-ray equipment.

Recall: Z-2108-2012 · Reported August 8, 2012

Enforcement

Recall Number
Z-2108-2012
Event ID
62590
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Maquet Medical Systems, Usa
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 8, 2012
Initiation Date
April 13, 2012
Classification Date
July 30, 2012
Termination Date
September 4, 2013
Address
45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States

Description

Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table top (1180-16XX) serves to support and position patients for surgical procedures (e.g. cardiovascular surgery) and interventional examination immediately before, during and after the surgical phase. The table top is radiotranslucent and enables intraoperative use of X-ray equipment.

Reason

As a result of a small number of customer complaints, MAQUET conducted an investigation and identified a potential problem with specific MAGNUS Carbon-Fibre Table Tops manufactured between February 23, 2010 and February 28, 2012 and distributed between November 11, 2010 and April 6, 2012. There is a potential issue with the motor brakes for the longitudinal shift drive, and the tilt drive.

Code Info

Model # 1180.16F2 Serial #s 1,3,7, & 8; Model # 1180.16F3 Serial # 1; Model #1180.16F5 Serial #4; Model # T285.7000 Serial #s 1,2,3,4, & 7.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of United Arab Emirates, Austria, Australia, Belgium, Brazil, Canada, Switzerland, China, Germany, Denmark, Spain, France, Hong Kong, Ireland, Israel, India, Italy, Japan, South Korea, Netherlands, Poland, Russia, Saudi Arabia, Sweden, Singapore, and Thailand.

Quantity

164 Units located in 27 Countries