Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142B; 6) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145; 7) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145B; 8) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5; 9) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5B; 10) NEURO PACK-LF, Model Number: DYNJ0305809O; 11) NEURO PACK-LF, Model Number: DYNJ0305809P; 12) NEURO PACK-LF, Model Number: DYNJ0408776Y; 13) NEURO PACK-LF, Model Number: DYNJ04596K; 14) NEURO/LAMINECTOMY BASIN PK-LF, Model Number: DYNJ0781823K; 15) BASIC NEURO PACK SMMC-LF, Model Number: DYNJ35243L; 16) NEURO PACK, Model Number: DYNJ49098I; 17) NEURO PACK, Model Number: DYNJ53089B; 18) NEURO LAMINECTOMY-LF, Model Number: DYNJ53799; 19) NEURO MINOR PACK-UNIV, Model Number: DYNJ58704A; 20) NEURO PACK - IFCAP 230316, Model Number: DYNJ60100C; 21) NEURO PACK, Model Number: DYNJ60792B; 22) NEURO PACK, Model Number: DYNJ60792C; 23) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157C; 24) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157D; 25) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157F; 26) CHRISTUS CHILDRENS NEURO PK, Model Number: DYNJ61168D; 27) NEURO PACK, Model Number: DYNJ61740D; 28) NEURO ST DAVIDS MEDICAL CTR, Model Number: DYNJ61998; 29) NEURO GOODIE PACK - DUNN, Model Number: DYNJ64570A; 30) DISCECTOMY PACK, Model Number: DYNJ67835; 31) FSC NEURO PACK, Model Number: DYNJ80442B; 32) WWD NEURO PACK, Model Number: DYNJ80849B; 33) WWD NEURO PACK, Model Number: DYNJ80849C; 34) WWD NEURO PACK, Model Number: DYNJ80849D; 35) WWD NEURO PACK, Model Number: DYNJ80849F; 36) WWD NEURO PACK, Model Number: DYNJ80849G; 37) WWD NEURO PACK, Model Number: DYNJ80849I; 38) WWD NEURO PACK, Model Number: DYNJ80849J; 39) NEURO DIAGNOSTIC, Model Number: DYNJ84653A; 40) NEURO, Model Number: DYNJ900904J; 41) NEURO, Model Number: DYNJ900904K; 42) NEURO, Model Number: DYNJ900904L; 43) NEURO, Model Number: DYNJ900904M; 44) NEURO, Model Number: DYNJ902585; 45) NEURO, Model Number: DYNJ902585A; 46) NEURO - IFCAP 230316, Model Number: DYNJ908393; 47) NEURO - IFCAP 230316, Model Number: DYNJ908393A; 48) NEURO - IFCAP 230316, Model Number: DYNJ908393B; 49) NEURO KIT, Model Number: DYNJ909573; 50) NEURO, Model Number: DYNJ910384; 51) NEURO, Model Number: DYNJ910384A; 52) NEURO PACK-LF, Model Number: PHS390015M
Enforcement
- Recall Number
- Z-2107-2026
- Event ID
- 98642
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 13, 2026
- Initiation Date
- January 7, 2026
- Classification Date
- May 7, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142B; 6) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145; 7) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145B; 8) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5; 9) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5B; 10) NEURO PACK-LF, Model Number: DYNJ0305809O; 11) NEURO PACK-LF, Model Number: DYNJ0305809P; 12) NEURO PACK-LF, Model Number: DYNJ0408776Y; 13) NEURO PACK-LF, Model Number: DYNJ04596K; 14) NEURO/LAMINECTOMY BASIN PK-LF, Model Number: DYNJ0781823K; 15) BASIC NEURO PACK SMMC-LF, Model Number: DYNJ35243L; 16) NEURO PACK, Model Number: DYNJ49098I; 17) NEURO PACK, Model Number: DYNJ53089B; 18) NEURO LAMINECTOMY-LF, Model Number: DYNJ53799; 19) NEURO MINOR PACK-UNIV, Model Number: DYNJ58704A; 20) NEURO PACK - IFCAP 230316, Model Number: DYNJ60100C; 21) NEURO PACK, Model Number: DYNJ60792B; 22) NEURO PACK, Model Number: DYNJ60792C; 23) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157C; 24) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157D; 25) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157F; 26) CHRISTUS CHILDRENS NEURO PK, Model Number: DYNJ61168D; 27) NEURO PACK, Model Number: DYNJ61740D; 28) NEURO ST DAVIDS MEDICAL CTR, Model Number: DYNJ61998; 29) NEURO GOODIE PACK - DUNN, Model Number: DYNJ64570A; 30) DISCECTOMY PACK, Model Number: DYNJ67835; 31) FSC NEURO PACK, Model Number: DYNJ80442B; 32) WWD NEURO PACK, Model Number: DYNJ80849B; 33) WWD NEURO PACK, Model Number: DYNJ80849C; 34) WWD NEURO PACK, Model Number: DYNJ80849D; 35) WWD NEURO PACK, Model Number: DYNJ80849F; 36) WWD NEURO PACK, Model Number: DYNJ80849G; 37) WWD NEURO PACK, Model Number: DYNJ80849I; 38) WWD NEURO PACK, Model Number: DYNJ80849J; 39) NEURO DIAGNOSTIC, Model Number: DYNJ84653A; 40) NEURO, Model Number: DYNJ900904J; 41) NEURO, Model Number: DYNJ900904K; 42) NEURO, Model Number: DYNJ900904L; 43) NEURO, Model Number: DYNJ900904M; 44) NEURO, Model Number: DYNJ902585; 45) NEURO, Model Number: DYNJ902585A; 46) NEURO - IFCAP 230316, Model Number: DYNJ908393; 47) NEURO - IFCAP 230316, Model Number: DYNJ908393A; 48) NEURO - IFCAP 230316, Model Number: DYNJ908393B; 49) NEURO KIT, Model Number: DYNJ909573; 50) NEURO, Model Number: DYNJ910384; 51) NEURO, Model Number: DYNJ910384A; 52) NEURO PACK-LF, Model Number: PHS390015M
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
1) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21CBA213; 2) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21DBD719; 3) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GMA403; 4) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GME436; 5) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GMF497; 6) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21JMH357; 7) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21KMD971; 8) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21LMG598; 9) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22DMD555; 10) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22HMC764; 11) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22KMF991; 12) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23AMJ086; 13) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23CMD444; 14) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23CMH970; 15) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23EME270; 16) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23GMC110; 17) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23IME897; 18) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23JMA572; 19) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23JMI786; 20) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23KMB758; 21) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 24CMA817; 22) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 24EMI161; 23) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 24IMH330; 24) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 24LMB440; 25) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 25CMC121; 26) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 25FMC026; 27) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21ABL131; 28) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21ABU817; 29) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21BBA521; 30) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21BBP552; 31) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21DBE120; 32) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21FBE311; 33) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21GBG761; 34) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21GBS468; 35) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21HBN643; 36) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21JBB951; 37) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21LBL164; 38) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22BBT015; 39) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22DBD203; 40) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22EBA074; 41) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22EBO542; 42) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22FBC967; 43) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22FBF303; 44) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22IBE871; 45) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22NBA962; 46) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22OBC508; 47) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21ABL138; 48) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21ABW870; 49) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21BBB221; 50) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21BBG852; 51) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22IBN422; 52) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22KBB452; 53) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22LBR466; 54) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23ABU154; 55) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23BBU041; 56) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23CBG764; 57) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23CBP083; 58) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23DBF392; 59) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23FBE493; 60) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23GBE625; 61) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23HBT876; 62) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23IBS183; 63) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23KBK072; 64) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24ABL221; 65) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24CBA173; 66) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24CBU410; 67) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24DBJ890; 68) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24EBQ164; 69) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24GBS072; 70) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24JBS893; 71) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24KBA532; 72) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24KBK044; 73) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24LBI083; 74) DYKMBNDL145, UDI-DI: 10193489478365(each), 40193489478366(case), Lot Number: 21ABT643; 75) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 22LBR481; 76) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 23EBU571; 77) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 23GBE626; 78) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 23GBS954; 79) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 23HBC122; 80) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 23HBH308; 81) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 23JBQ486; 82) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 23LBH331; 83) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 24ABH120; 84) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 24ABT109; 85) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 24HBF314; 86) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 24KBB216; 87) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 24LBL113; 88) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 25BBG293; 89) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 25CBS910; 90) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 25DBI287; 91) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 25FBH445; 92) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 25GBF152; 93) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 25GBF317; 94) DYKMBNDL5, UDI-DI: 10193489477634(each), 40193489477635(case), Lot Number: 21ABQ194; 95) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 22IBJ928; 96) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23ABK051; 97) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23EBB597; 98) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23FBP269; 99) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23LBL844; 100) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24BBG003; 101) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24DBJ607; 102) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24EBC897; 103) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24GBK951; 104) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24LBH040; 105) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24LBJ842; 106) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25BBM404; 107) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25DBO558; 108) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25FBD172; 109) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25HBO105; 110) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25KBQ729; 111) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22HBF833; 112) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22HBO559; 113) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22JBL998; 114) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22LBL870; 115) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22OBL140; 116) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23CBI776; 117) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23DBQ920; 118) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23EBW271; 119) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23GBD741; 120) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23IBR967; 121) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 23LBR942; 122) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 24ABW618; 123) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 24CBL727; 124) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 24DBT971; 125) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 24FMG901; 126) DYNJ0408776Y, UDI-DI: 10195327115852(each), 40195327115853(case), Lot Number: 23DBS739; 127) DYNJ04596K, UDI-DI: 10193489301342(each), 40193489301343(case), Lot Number: 23KMD949; 128) DYNJ0781823K, UDI-DI: 10193489919196(each), 40193489919197(case), Lot Number: 23IME718; 129) DYNJ35243L, UDI-DI: 10195327638733(each), 40195327638734(case), Lot Number: 24FMB808; 130) DYNJ49098I, UDI-DI: 10193489712100(each), 40193489712101(case), Lot Number: 21BBP868; 131) DYNJ53089B, UDI-DI: 10193489335248(each), 40193489335249(case), Lot Number: 21BBA722; 132) DYNJ53799, UDI-DI: 10889942444557(each), 40889942444558(case), Lot Number: 21ADA264; 133) DYNJ58704A, UDI-DI: 10193489502053(each), 40193489502054(case), Lot Number: 23CBT449; 134) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21GLB218; 135) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21HLA233; 136) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21HLA398; 137) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21KLA157; 138) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21EBI340; 139) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21CBC497; 140) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21DBJ319; 141) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21DBO518; 142) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21EBR912; 143) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21EBR913; 144) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21FBS847; 145) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 22ABJ128; 146) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 22HBN525; 147) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 22OBE182; 148) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23CBE668; 149) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23DBB793; 150) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23EBM305; 151) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23HBY836; 152) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23KBH242; 153) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23LBN124; 154) DYNJ61157C, UDI-DI: 10195327470234(each), 40195327470235(case), Lot Number: 23IBP298; 155) DYNJ61157C, UDI-DI: 10195327470234(each), 40195327470235(case), Lot Number: 23KBP559; 156) DYNJ61157C, UDI-DI: 10195327470234(each), 40195327470235(case), Lot Number: 23LBG606; 157) DYNJ61157D, UDI-DI: 10195327605919(each), 40195327605910(case), Lot Number: 24ABQ754; 158) DYNJ61157D, UDI-DI: 10195327605919(each), 40195327605910(case), Lot Number: 24BBH233; 159) DYNJ61157F, UDI-DI: 10195327690984(each), 40195327690985(case), Lot Number: 24EBO693; 160) DYNJ61157F, UDI-DI: 10195327690984(each), 40195327690985(case), Lot Number: 24GBA775; 161) DYNJ61157F, UDI-DI: 10195327690984(each), 40195327690985(case), Lot Number: 24IBA986; 162) DYNJ61157F, UDI-DI: 10195327690984(each), 40195327690985(case), Lot Number: 24IBQ450; 163) DYNJ61157F, UDI-DI: 10195327690984(each), 40195327690985(case), Lot Number: 24KBM686; 164) DYNJ61157F, UDI-DI: 10195327690984(each), 40195327690985(case), Lot Number: 24LBL683; 165) DYNJ61157F, UDI-DI: 10195327690984(each), 40195327690985(case), Lot Number: 25CBK270; 166) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24CBB175; 167) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24GBA833; 168) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24IBQ689; 169) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25ABG065; 170) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25CBH417; 171) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25EBE757; 172) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25FBF194; 173) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25GBR898; 174) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25HBG888; 175) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25HBR218; 176) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25KBL642; 177) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21DBV445; 178) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21KBA371; 179) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBB713; 180) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBB714; 181) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBE011; 182) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22ABL607; 183) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22CBP375; 184) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22CBX534; 185) DYNJ61998, UDI-DI: 10193489615012(each), 40193489615013(case), Lot Number: 23EBV072; 186) DYNJ64570A, UDI-DI: 10195327242800(each), 40195327242801(case), Lot Number: 22LBK925; 187) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22EBL897; 188) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22GBR040; 189) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22IBV281; 190) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22LBK279; 191) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23BBS499; 192) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23CBR140; 193) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23CBT224; 194) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23DBP210; 195) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23FBJ662; 196) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23GBG567; 197) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 24CBB478; 198) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 24GBQ829; 199) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 24IBL709; 200) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 24JBR686; 201) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 24LBR882; 202) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 25BBD822; 203) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 25BBH829; 204) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 25EBR504; 205) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 25FBO441; 206) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 25HBO015; 207) DYNJ80442B, UDI-DI: 10195327516338(each), 40195327516339(case), Lot Number: 25HBS113; 208) DYNJ80849B, UDI-DI: 10195327443689(each), 40195327443680(case), Lot Number: 23GBF525; 209) DYNJ80849B, UDI-DI: 10195327443689(each), 40195327443680(case), Lot Number: 23HBR538; 210) DYNJ80849C, UDI-DI: 10195327518561(each), 40195327518562(case), Lot Number: 23JBE317; 211) DYNJ80849C, UDI-DI: 10195327518561(each), 40195327518562(case), Lot Number: 23KBI537; 212) DYNJ80849D, UDI-DI: 10195327579289(each), 40195327579280(case), Lot Number: 24ABE404; 213) DYNJ80849D, UDI-DI: 10195327579289(each), 40195327579280(case), Lot Number: 24CBC623; 214) DYNJ80849D, UDI-DI: 10195327579289(each), 40195327579280(case), Lot Number: 24DBQ059; 215) DYNJ80849F, UDI-DI: 10198459032165(each), 40198459032166(case), Lot Number: 24GBO690; 216) DYNJ80849F, UDI-DI: 10198459032165(each), 40198459032166(case), Lot Number: 24HBM361; 217) DYNJ80849F, UDI-DI: 10198459032165(each), 40198459032166(case), Lot Number: 24JBJ479; 218) DYNJ80849G, UDI-DI: 10198459166129(each), 40198459166120(case), Lot Number: 24LBK924; 219) DYNJ80849I, UDI-DI: 10198459233159(each), 40198459233150(case), Lot Number: 25ABS617; 220) DYNJ80849I, UDI-DI: 10198459233159(each), 40198459233150(case), Lot Number: 25CBQ272; 221) DYNJ80849I, UDI-DI: 10198459233159(each), 40198459233150(case), Lot Number: 25GBE969; 222) DYNJ80849J, UDI-DI: 10198459561245(each), 40198459561246(case), Lot Number: 25KBJ851; 223) DYNJ84653A, UDI-DI: 10195327697419(each), 40195327697410(case), Lot Number: 24GBU616; 224) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21GBF913; 225) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21HBI172; 226) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21JBQ824; 227) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21LBG281; 228) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22CLA766; 229) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22EBB675; 230) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22ELA806; 231) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBJ991; 232) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBJ992; 233) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBR893; 234) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22JBY355; 235) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22KBC323; 236) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22OBE679; 237) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23ABC415; 238) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23ABR127; 239) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23FBN975; 240) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23FBV444; 241) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23GBW505; 242) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23HBQ501; 243) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23JBF549; 244) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23KBH986; 245) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24ABH283; 246) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24CBN849; 247) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24DMB555; 248) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24DMH051; 249) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24FMF700; 250) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24HMA754; 251) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24JMB004; 252) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24JMF168; 253) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24KMG047; 254) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24LMD278; 255) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 25AMG870; 256) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 25BME471; 257) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25CMJ212; 258) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25DME524; 259) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25EMG513; 260) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25EMJ902; 261) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25FMG005; 262) DYNJ900904M, UDI-DI: 10198459408519(each), 40198459408510(case), Lot Number: 25GMI512; 263) DYNJ900904M, UDI-DI: 10198459408519(each), 40198459408510(case), Lot Number: 25HMI564; 264) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21CDB697; 265) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21EDA724; 266) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21FDA224; 267) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21GDB745; 268) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21IDB884; 269) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21JDC066; 270) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21LDA963; 271) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22FDB376; 272) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22GDB053; 273) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22HDB311; 274) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22IDA639; 275) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23ADA425; 276) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23FDA576; 277) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23IDB360; 278) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24CDA135; 279) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24GDA318; 280) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24JDC387; 281) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24LDA154; 282) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 25ADA139; 283) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 25DDB556; 284) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22BDB502; 285) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22EDA370; 286) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22GDA427; 287) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22GDB437; 288) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22HDC481; 289) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22IDA609; 290) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22IDB239; 291) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22LDA043; 292) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23BDB148; 293) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23EDB018; 294) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23FDB736; 295) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23GDA913; 296) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23HDC123; 297) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23HDC701; 298) DYNJ908393A, UDI-DI: 10195327490416(each), 40195327490417(case), Lot Number: 23KDA234; 299) DYNJ908393A, UDI-DI: 10195327490416(each), 40195327490417(case), Lot Number: 24BDA090; 300) DYNJ908393A, UDI-DI: 10195327490416(each), 40195327490417(case), Lot Number: 24CDB967; 301) DYNJ908393A, UDI-DI: 10195327490416(each), 40195327490417(case), Lot Number: 24EDB090; 302) DYNJ908393A, UDI-DI: 10195327490416(each), 40195327490417(case), Lot Number: 24EDC063; 303) DYNJ908393A, UDI-DI: 10195327490416(each), 40195327490417(case), Lot Number: 24IDA719; 304) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 24JDA510; 305) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 24KDB904; 306) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 25CDA661; 307) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 25DDC013; 308) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 25EDA457; 309) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 25FDB472; 310) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 25HDA240; 311) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 25LDA938; 312) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 25KDB530; 313) DYNJ908393B, UDI-DI: 10198459118920(each), 40198459118921(case), Lot Number: 25IDB199; 314) DYNJ909573, UDI-DI: 10195327404741(each), 40195327404742(case), Lot Number: 25DLA489; 315) DYNJ909573, UDI-DI: 10195327404741(each), 40195327404742(case), Lot Number: 25FLA327; 316) DYNJ910384, UDI-DI: 10195327661595(each), 40195327661596(case), Lot Number: 24EBH096; 317) DYNJ910384, UDI-DI: 10195327661595(each), 40195327661596(case), Lot Number: 24FBF785; 318) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 24JBQ224; 319) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25ABA025; 320) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25BBN770; 321) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25CBF584; 322) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25CBF621; 323) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25DBC086; 324) PHS390015M, UDI-DI: 10193489333824(each), 40193489333825(case), Lot Number: 20LMD456
Worldwide - US Nationwide distribution.
10184 units