FDA Enforcement Class II Ongoing

VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.

Recall: Z-2105-2025 · Reported July 16, 2025

Enforcement

Recall Number
Z-2105-2025
Event ID
97036
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
July 16, 2025
Initiation Date
June 10, 2025
Classification Date
July 10, 2025
Address
4100 Hamline Ave N, N/A, Saint Paul, MN, 55112-5700, United States

Description

VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.

Reason

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

Code Info

VXS0201 UDI-DI 00685447020301 Lot 35800622; VXS2112 UDI-DI 00685447019251 Lot 35892045

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.

Quantity

14,724