FDA Enforcement Class II Terminated

YXLON Cheetah cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control

Recall: Z-2105-2019 · Reported August 14, 2019

Enforcement

Recall Number
Z-2105-2019
Event ID
83295
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Comet Technologies Usa Inc
Voluntary / Mandated
FDA Mandated
Initial Notification
Letter
Report Date
August 14, 2019
Initiation Date
October 1, 2018
Classification Date
August 7, 2019
Termination Date
June 13, 2024
Address
100 Trap Falls Rd, N/A, Shelton, CT, 06484-4604, United States

Description

YXLON Cheetah cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control

Reason

Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.

Code Info

X-Ray Systems

Distribution

US Nationwide Distribution

Quantity

77