FDA Enforcement Class I Ongoing

The Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009

Recall: Z-2104-2023 · Reported July 19, 2023

Enforcement

Recall Number
Z-2104-2023
Event ID
92562
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 19, 2023
Initiation Date
June 15, 2023
Classification Date
July 13, 2023
Address
1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States

Description

The Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009

Reason

A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01.

Code Info

UDI/DI 00085412610900, Software Version v9.02.01, All Serial Numbers

Distribution

US Nationwide distribution.

Quantity

19861 units