Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
Enforcement
- Recall Number
- Z-2100-2017
- Event ID
- 77177
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Mobius Medical Systems, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 24, 2017
- Initiation Date
- February 23, 2015
- Classification Date
- May 16, 2017
- Termination Date
- July 10, 2017
- Address
- 4615 Southwest Fwy Ste 330, Houston, TX, 77027-7106, United States
Description
Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
Mobius3D version 1.5.0 contained a defect in software code which affects users who perform beam customization, and may lead to a discrepancy in dose calculation between this version of Mobius3D (1.5.0) and the prior version (1.4.2), where none should have been expected.
Version 1.5.0 and 1.5.2 of the Mobius3D software is affected.
Worldwide distribution - US Nationwide and the countries of Sweden, Canada, Netherlands, Denmark, Spain, Switzerland, Australia, Austria, Israel, United Kingdom, France