FDA Enforcement
Class II
Terminated
NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
Recall: Z-2099-2021
·
Reported July 28, 2021
Enforcement
- Recall Number
- Z-2099-2021
- Event ID
- 88130
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Randox Laboratories Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 28, 2021
- Initiation Date
- May 10, 2021
- Classification Date
- July 20, 2021
- Termination Date
- November 10, 2021
- Address
- 55 Diamond Road, N/A, Crumlin (North), N/A, N/A, Ireland
Description
NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
Reason
Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples
Code Info
Batch Number: 544642 Exp. Date: 28th Nov 2022 GTIN: 05055273203066
Distribution
US Nationwide Distribution in the states of AR, CA, CT, NY, OH, PA, TN, and TX
Quantity
71 kits US