FDA Enforcement Class II Ongoing

Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Recall: Z-2098-2025 · Reported July 16, 2025

Enforcement

Recall Number
Z-2098-2025
Event ID
97193
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 16, 2025
Initiation Date
June 20, 2025
Classification Date
July 8, 2025
Address
4, Hayozma St, N/A, Tirat Carmel, N/A, N/A, Israel

Description

Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Reason

Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.

Code Info

Serial Numbers: Millennium VG 1074; "Varicam Millennium VG Discovery VH" 1003 1092 10147 10242 10374 10378 10539 10543 10550 10553 10561 10569 10570 10571 10579 10582 10587 10599 10604 3101-1502

Distribution

Worldwide - US Nationwide distribution.

Quantity

21