FDA Enforcement Class II Terminated

Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.

Recall: Z-2097-2015 · Reported July 29, 2015

Enforcement

Recall Number
Z-2097-2015
Event ID
71446
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Maquet Cardiovascular Us Sales, Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 29, 2015
Initiation Date
June 10, 2015
Classification Date
July 17, 2015
Termination Date
April 18, 2016
Address
45 Barbour Pond Drive, N/A, Wayne, NJ, 07470, United States

Description

Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.

Reason

There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.

Code Info

SN 00048

Distribution

Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwan and South Africa.

Quantity

12 units