FDA Enforcement
Class II
Terminated
AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and Irrigation Sleeve (Model Number NGP0020) Used with the AMO Whitestar Signature System. For ophthalmic use.
Recall: Z-2096-2017
·
Reported May 24, 2017
Enforcement
- Recall Number
- Z-2096-2017
- Event ID
- 77024
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Medical Optics Inc. (AMO)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 24, 2017
- Initiation Date
- March 30, 2017
- Classification Date
- May 16, 2017
- Termination Date
- August 23, 2017
- Address
- 1700 E Saint Andrew Pl, Santa Ana, CA, 92705-4933, United States
Description
AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and Irrigation Sleeve (Model Number NGP0020) Used with the AMO Whitestar Signature System. For ophthalmic use.
Reason
Packaging error. 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package.
Code Info
Part Number NGP0025 Lot CA06182 (72 units) Part Number NGP0020 Lot CA06180 (288 units)
Distribution
Worldwide distribution to SA, TR, AZ, NL, PT, SG, AU, JP, and HK.
Quantity
360 units total