Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and operating room. Absorber Adapter Assembly Service Kit, part number: 050-9045-00. Absorber Adapter Assembly, part number: 650-1648-01. Arkon Absorber Canister Carrier, part number: 380-0916-01.
Enforcement
- Recall Number
- Z-2093-2017
- Event ID
- 77019
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Del Mar Reynolds Medical, Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 24, 2017
- Initiation Date
- April 14, 2017
- Classification Date
- May 16, 2017
- Termination Date
- October 16, 2017
- Address
- 1 2 Harforde Court John Tate Road, Business Park, Hertford, N/A, United Kingdom
Description
Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and operating room. Absorber Adapter Assembly Service Kit, part number: 050-9045-00. Absorber Adapter Assembly, part number: 650-1648-01. Arkon Absorber Canister Carrier, part number: 380-0916-01.
Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.
*********the serial numbers and part numbers of units distributed in the US ********* Absorber Adapter Assembly Service Kit part number: 050-9045-00. Devices serial numbers: ARKN-000395, ARKN-000396, ARKN-000397, ARKN-000398, ARKN-000399, ARKN-000400, ARKN-000401, ARKN-000402, ARKN-000403, ARKN-000404, ARKN-000405, ARKN-000406, ARKN-000407, ARKN-000408, ARKN-000409, ARKN-000410, ARKN-000411, ARKN-000412, and ARKN-000413. *********the serial numbers of units distributed international ************************** ARKN-000375, ARKN-000414, ARKN-000415, ARKN-000416, ARKN-000417, ARKN-000418, ARKN-000419, ARKN-000420, ARKN-000421, ARKN-000422, and ARKN-000423.
US distribution in FL, LA, MS, NC, NC, SC and WY and Internationally to Panama and United Kingdom.
62 total (US= 19 devices and 32 kits; International = 11 devices)