NDI Disposable Reflective Marker Spheres for Brainlab IGS (Image Guided Surgery) system. Product Usage: Disposable reflective marker spheres are attached to reference arrays and instruments, thus enabling the infrared tracking systems to detect the position of the patient and instruments in the surgical field.

Recall: Z-2093-2015 · Reported July 22, 2015

Enforcement

Recall Number
Z-2093-2015
Event ID
71350
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Northern Digital Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 22, 2015
Initiation Date
May 20, 2015
Classification Date
July 16, 2015
Termination Date
October 13, 2017
Address
103 Randall Drive, N/A, Waterloo, Ontario, N/A, Canada

Description

NDI Disposable Reflective Marker Spheres for Brainlab IGS (Image Guided Surgery) system. Product Usage: Disposable reflective marker spheres are attached to reference arrays and instruments, thus enabling the infrared tracking systems to detect the position of the patient and instruments in the surgical field.

Reason

Inadequate curing of adhesive, which may cause the two halves of the spheres to separate.

Code Info

1) Catalog Number: 41773 (90 pack) Lot # C102061501

Distribution

Worldwide - US Nationwide Distribution and the countries of Australia and Canada.

Quantity

32,400 spheres