FDA Enforcement Class II Terminated

Solanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior Stabilization System, a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).

Recall: Z-2092-2014 · Reported July 30, 2014

Enforcement

Recall Number
Z-2092-2014
Event ID
68667
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Alphatec Spine, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 30, 2014
Initiation Date
June 26, 2014
Classification Date
July 22, 2014
Termination Date
September 3, 2014
Address
5818 El Camino Real, Carlsbad, CA, 92008-8816, United States

Description

Solanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior Stabilization System, a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).

Reason

Alphatec Spine is recalling the 2.1mm Fixed Depth Drill 10mm because they received a report that the instrument malfunctioned and golden stop slid from its fixed position during use.

Code Info

Lot No. 7117902

Distribution

Nationwide Distribution including CT, NC, AL, WV, IL, AR, and NY.

Quantity

14 units total (12 units in US)