FDA Enforcement Class I Ongoing

Brand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

Recall: Z-2085-2025 · Reported July 16, 2025

Enforcement

Recall Number
Z-2085-2025
Event ID
96943
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Edwards Lifesciences, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 16, 2025
Initiation Date
May 14, 2025
Classification Date
July 10, 2025
Address
1 Edwards Way, N/A, Irvine, CA, 92614-5688, United States

Description

Brand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

Reason

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

Code Info

Lot Code: Model No: FEMII018AS; UDI: 00690103168358; All lots up to: BSLC9064 Serial Numbers: 131776 BSLC4190 BSLC9064

Distribution

Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates (UAE), United Kingdom.

Quantity

558