FDA Enforcement Class II Terminated

ST-AIA DHEA-S; Part Number: 025222 Assay, Reproductive Hormone

Recall: Z-2085-2018 · Reported June 13, 2018

Enforcement

Recall Number
Z-2085-2018
Event ID
79813
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Tosoh Bioscience Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 13, 2018
Initiation Date
March 5, 2018
Classification Date
June 5, 2018
Termination Date
August 31, 2020
Address
3600 Gantz Rd, N/A, Grove City, OH, 43123-1895, United States

Description

ST-AIA DHEA-S; Part Number: 025222 Assay, Reproductive Hormone

Reason

Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.

Code Info

All Lot Codes

Distribution

Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.

Quantity

1442