FDA Enforcement Class II Terminated

NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number 989605858501. The device is labeled in part: "***Neusoft***NeuViz16 MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED FOR NEUSOFT MEDICAL SYSTEMS CO., LTD***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD***ADDRESS: NEUSOFT PARK, HUNNAN INDUSTRIAL AREA, SHENYANG***CHINA***".

Recall: Z-2083-2012 · Reported August 1, 2012

Enforcement

Recall Number
Z-2083-2012
Event ID
62481
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips And Neusoft Medical Systems Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 1, 2012
Initiation Date
June 8, 2012
Classification Date
July 26, 2012
Termination Date
September 17, 2012
Address
16 Century Road, Neusoft Park,, Hun Nan Industrial Area, Shenyang, Liaoning, N/A, China

Description

NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number 989605858501. The device is labeled in part: "***Neusoft***NeuViz16 MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED FOR NEUSOFT MEDICAL SYSTEMS CO., LTD***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD***ADDRESS: NEUSOFT PARK, HUNNAN INDUSTRIAL AREA, SHENYANG***CHINA***".

Reason

One of the four screws that secure the Oil-accumulator to the X-Ray Tube pulled out in the NeuViz 16 CT scanner, which caused the Oil-accumulator to detach.

Code Info

Serial No.: N16E0 90002, N16E0 90003, N16E0 90004, N16E0 90007, N16E0 90008, N16E0 90010, N16E1 00015, N16E1 00017, N16E1 00023, N16E1 10016, N16E1 10023, N16E1 10026, N16E1 10027, N16E1 10040, and N16E1 10053.

Distribution

Nationwide Distribution -- North Carolina, Ohio, and Texas.

Quantity

15 units distributed in the US