FDA Enforcement Class I Ongoing

Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

Recall: Z-2081-2025 · Reported July 16, 2025

Enforcement

Recall Number
Z-2081-2025
Event ID
96943
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Edwards Lifesciences, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 16, 2025
Initiation Date
May 14, 2025
Classification Date
July 10, 2025
Address
1 Edwards Way, N/A, Irvine, CA, 92614-5688, United States

Description

Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

Reason

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

Code Info

Lot Code: Model No: OPTI18; UDI: 00690103180565; All lots up to: BSLC9155 Serial Numbers: 130451 130466 130472 130593 130709 130764 130818 130821 130847 130948 130952 131038 131043 131047 131182 131193 131231 131236 131289 131311 131317 131433 131435 131513 131516 131553 131554 131606 131623 131625 131628 131682 131737 131909 131929 131966 132175 132201 132366 132395 132416 132430 132435 132505 132832 132958 132959 BSLC0110 BSLC0166 BSLC0661 BSLC0661 BSLC0826 BSLC0826 BSLC0826 BSLC0826 BSLC0855 BSLC0855 BSLC0887 BSLC0887 BSLC0979 BSLC1114 BSLC1114 BSLC1211 BSLC1308 BSLC1469 BSLC1656 BSLC1660 BSLC1815 BSLC3221 BSLC4153 BSLC4400 BSLC4461 BSLC4513 BSLC5027 BSLC5027 BSLC5320 BSLC5367 BSLC5765 BSLC6421 BSLC6467 BSLC7335 BSLC7514 BSLC7515 BSLC8175 BSLC8385 BSLC8535 BSLC8657 BSLC9100 BSLC9104 BSLC9105 BSLC9152 BSLC9155

Distribution

Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates (UAE), United Kingdom.

Quantity

13240